Rev. 6.4 dated 2022-02
Software Released Version: 1
Part 1 • Page 13-10
Menu selection
Part 1 – Operating Instructions
(Eco)RO Dia II
Part 1 • Chap
ter 13
13.11 Operating hours meter, menu item 11
13.12 Hot RO, menu item 12
Menu item 12 is only applicable to reverse osmosis systems with the Hot RO option.
13.13 Service mode, menu item 13
Op. hours meter
Pump M1 ######## h
Pump M2 ######## h
Pump M3 ######## h
System ######## h
MV Y2 #########
MV Y3 #########
MV Y4 #########
MV Y5 #########
MV Y6 #########
MV Y7 #########
MV Y8 #########
MV Y9 #########
MV Y10 #########
MV Y30 #########
MV Y5.1.1 #########
<<
The most important operating hours (depending on the type set) can be
viewed in the operating hours meter menu item.
Pump M1
Pump M2
Pump M3
System
MV Y2
MV Y3
MV Y4
MV Y5
MV Y6
MV Y7
MV Y8
MV Y9
MV Y10
MV Y30
MV Y5.1.1
Upstream of 1
st
osmosis membrane stage
Upstream of 1
st
osmosis membrane stage
Upstream of 2
nd
osmosis membrane stage
Reverse osmosis system
Concentrate return, 1
st
stage
Impulse rinsing, 1
st
stage
Soft water feed upstream of the 1
st
stage
Permeate feed in ring piping
Rinsing 1
st
stage
Rinsing 2
nd
stage
Concentrate return, 2
nd
stage
Concentrate discharge in channel
Soft water feed in tank
Impulse shear force rinsing (ISS) reserve
Ring piping return reserve
13 Service mode
Digital inputs
Digital outputs
<<
OK
Digital inputs and outputs can be imported or set in service mode. This is
used to check the functional ability of the components.
Digital inputs
Function (“1”, if ...)
Dialysis mode
Selection switch to (position 1) or “2” + “program start, dialysis”
Standby mode
Selection switch to 2 + program step standby mode (stby)
Ctrl volt. cond.+SL
External conductivity measuring instrument
Ctrl volt. RO I
Ctrl volt. RO II
Pump M1
Pump OK
Pump M2
Pump OK