Rev. 6.4 dated 2022-02
Software Released Version: 1
Part 1 • Page 7-2
Description of components
Part 1 – Operating Instructions
(Eco)RO Dia II
Part 1 • Chapter 7
7.2
Signal lamps
•
“
Dialysis mode
” lamp lights up
The system is in dialysis mode
•
“
Standby mode
” lamp lights up
The system is in standby mode
•
“
Rinsing
” lamp lights up
When IRS 1 is active
•
“
Emergency mode
” lamp lights up
The control unit has failed. The system continues to product the full amount, however, since the
power supply has not failed. In this switching state, rinsing no longer occurs. Please contact the B.
Braun service department.
•
“
Alarm
” lamp lights up
As soon as the fault lamp lights up, the type of fault appears on the display. Each fault is displayed
with a comment. Please locate the cause according to the process diagram
Additional fault causes with comments are described in the text. During emergency mode
chapter 14 switch flap valves
K1
and
K3
according to the operating mode.
NOTE
If there is a power failure, operation is interrupted. This means shutdown occurs without
rinsing. Once power has been restored, the system starts with the operating mode it was in
when it switched off.