Maximum peak output voltage (up)
Note
The following graph (see Fig. 3) allows the user to judge whether the HF
device or its output setting is suitable for a given accessory (insulation rat-
ing).
Fig. 3
Maximum peak output voltage (up)
3.
Preparation and setup
Non-compliance with the following instructions will preclude all respon-
sibility and liability in this respect on the part of Aesculap.
►
When setting up and operating the product, adhere to
– national regulations for installation and operation,
– national regulations on fire and explosion protection.
Note
For the safety of patients and users it is essential that the mains power cord
and, especially, the protective earth connection are intact. In many cases
defective or missing protective earth connections are not registered imme-
diately.
►
Connect the device via the potential equalization terminal at the rear
panel of the device to the potential equalization system of the room
used for medical purposes.
Note
The potential equalization lead can be ordered from the manufacturer as
art. no. GK535 (4 m length) or TA008205 (0.8 m length).
3.1
First use
3.2
Stacking of units
The stacking provisions integrated in the device cover allows secure
mounting of an accessory device (such as the JET irrigation unit GN090)
on top of the HF device GN060.
►
Make certain the system is set up on a sufficiently stable support (e.g.
a table, ceiling support, equipment cart, etc.).
Stacking products fitted with stacking devices:
Fig. 4
Stacking of units
Legend
A
Accessory device, e.g. JET irrigation unit GN090
B
Locking screw
C
Stacking cone
►
Do not exceed a stack height of 475 mm.
►
Remove the caps from stacking cones
C
.
►
Unscrew the feet of the product that will be on top.
►
Move stacking cones
C
to the correct mounting position by applying a
screwdriver on locking screw
B
and turning it counterclockwise to the
positive stop.
►
Place the upper device on the one below it.
►
Securing the device: Turn locking screw
B
clockwise to the positive
stop.
►
Slightly lift the units to check that they are safely connected to each
other.
►
Place the units in a stable position.
WARNING
Risk of injury to patients or users caused by inade-
quate accessories (insulation rating)!
►
Make certain that the accessory voltage rating
specified in the product documentation is
higher than maximum peak output voltage for
the intended setting.
0
50
100
150
200
250
300
350
0 10 20 30 40 50
Micro
Macro
Output Set
ƟŶŐ
U
out
[V
]
WARNING
Risk of injury and/or product malfunction due to
incorrect operation of the electromedical system!
►
Adhere to the instructions for use of any medi-
cal device.
A
C
B
Содержание Aesculap GN060
Страница 2: ......
Страница 3: ...1 2 3 9 5 4 6 7 8 12 12 10 13 11 14 15 14 16 18 19 12 12 20 13 13 17 21 22 23...
Страница 150: ...148 Aesculap GN060 1 GN060 TA022130 Aesculap IEC VDE 1 1 IEC VDE...
Страница 151: ...149...
Страница 154: ...152 Aesculap GN060 3 Aesculap GK535 4 TA008205 0 8 3 1 3 2 GN060 JET GN090 4 A JET GN090 B C 475 C C B B 3 3 A C B...
Страница 155: ...153 3 4 5 B B 3 5 230 B A 25 24 C 25 6 23 24 25 A B C B C 100 120 10 110 220 240 10 230 23 A 25 24 C B...
Страница 159: ...157 F2 30 F3 30 F3 20 20 10 4 5 CPU RAM EPROM 10 RAM RAM EEPROM...
Страница 161: ...159 5 3 5 4 5 5 I T C F t I 1 Meliseptol HBV 50 1...
Страница 162: ...160 Aesculap GN060 5 6 5 7 6 GN060 6 1 B Braun Aesculap 10 C 3...
Страница 167: ...165 10 1 700 3000 11 B Braun Aesculap Extranet PDF...