Working with the HF surgical device
GN060
4.1
System set-up
Connecting the accessories
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
All configurations must comply with basic standard IEC/DIN EN 60601-1.
Any individual connecting devices with one another is responsible for such
configuration and must ensure compliance with basic standard IEC/DIN
EN 60601-1 or applicable national standards.
The insulation of the accessories (e.g. HF cables, instruments) must be
adequately rated for the maximum peak output voltage (see IEC/DIN
EN 60601-2-2).
►
Please contact your B. Braun/Aesculap partner or Aesculap Technical
Service (address: see Technical Service) with any inquiries in this
respect.
Connecting the power supply
Note
Before operating the device, make certain that the supply voltage marked
on the plug of the device matches that of the supply voltage being used. To
change the setting, see Presetting the correct voltage.
The HF device GN060 must only be operated with an alternating current
from 50 Hz to 60 Hz. The mains voltage must be between 100 V and 120 V
or between 220 V and 240 V. The voltage is preselected on the mains volt-
age connector on the rear side of the device. Insert the power cord only in
a properly mounted schuko power socket (for this see VDE 0107).
To prepare the device for operation, proceed as follows:
►
Connect the HF device to the mains voltage using a power cord (LED
1
illuminates).
►
Connect the bipolar cable to the socket
14
or
15
.
►
Connect foot switch to the socket
16
.
►
Activate button
3
(LED
1
and LED
2
illuminate).
The HF device is ready for operation.
Note
The isolation of the power cable from the device must also be treated as
secure isolation from the mains in the sense of the IEC norm. When aligning
the device, ensure that an isolation of the mains plug is possible without
any problems.
4.2
Function checks
►
Connect the HF device to the mains voltage.
Green LED
1
must illuminate.
►
Activate "Device on/off" button
3
.
LED
2
must illuminate. All LEDs and segments of the display illuminate
briefly.
For HF devices with the "automatic starting" option: In the display, the
"AU" display appears briefly.
The HF device is operable based on the basic settings approximately 1 s
after activation of the button.
The LEDs
5
and
18
illuminate in the display field.
In the display, "0.1" must be displayed as a dosage level.
►
Press "+" key
10
.
The dosage level must be able to be increased to 9.9.
►
Press "-" key
7
.
The dosage level must be able to be reduced to 0.1.
►
Press key
4
.
For signaling the "Macro" mode, LED
6
in the display field must illumi-
nate. "01" must be displayed as a dosage level.
The dosage level can be set between the limit levels "01" and "50" with
the "+" button
10
and the "-" button
7
.
►
If the "automatic starting" option is built in: Press key
19
.
The activation type must change. The activation type is displayed with
LED
17
or LED
18
.
►
Check memory function see Store device settings and see Retrieve
stored device settings.
►
Connect foot switch to the socket
16
and activate.
The acoustic and visual operating display
9
must correspond.
DANGER
Risk of injury due to unapproved configuration
using additional components!
►
For all applied components, ensure that their
classification matches that of the application
component (e.g. Type BF or Type CF) of the
respective device.
DANGER
Risk to life for patients and users due to inadver-
tent current supply to an instrument!
►
Make certain that not more than one patient
cable is connected to any output socket.
DANGER
Risk of death by electric shock!
►
Only ever connect the product to power mains
with equipment grounding conductor.
►
Set up the device in such a way that disconnect-
ing the power cord is straightforwardly possible.
Содержание Aesculap GN060
Страница 2: ......
Страница 3: ...1 2 3 9 5 4 6 7 8 12 12 10 13 11 14 15 14 16 18 19 12 12 20 13 13 17 21 22 23...
Страница 150: ...148 Aesculap GN060 1 GN060 TA022130 Aesculap IEC VDE 1 1 IEC VDE...
Страница 151: ...149...
Страница 154: ...152 Aesculap GN060 3 Aesculap GK535 4 TA008205 0 8 3 1 3 2 GN060 JET GN090 4 A JET GN090 B C 475 C C B B 3 3 A C B...
Страница 155: ...153 3 4 5 B B 3 5 230 B A 25 24 C 25 6 23 24 25 A B C B C 100 120 10 110 220 240 10 230 23 A 25 24 C B...
Страница 159: ...157 F2 30 F3 30 F3 20 20 10 4 5 CPU RAM EPROM 10 RAM RAM EEPROM...
Страница 161: ...159 5 3 5 4 5 5 I T C F t I 1 Meliseptol HBV 50 1...
Страница 162: ...160 Aesculap GN060 5 6 5 7 6 GN060 6 1 B Braun Aesculap 10 C 3...
Страница 167: ...165 10 1 700 3000 11 B Braun Aesculap Extranet PDF...