17
When trouble is found, check the following again before contacting Brasseler USA. If none of these is applicable or the trouble
is not remedied even after action has been taken, a failure of this product is suspected. Contact Brasseler USA.
<Control Unit, AC Adapter>
Problem
Cause
Solution
11. TROUBLESHOOTING
<Motor Handpiece>
Problem
Cause
Solution
The power is not turned on.
AC IN lamp does not light.
The battery charger does
not work.
(The CHARGE lamp does
not light.)
The AC adapter is not connected.
The plug of the AC adapter is not inserted
into the outlet, or there is no electricity in the
outlet.
The batteries are empty.
No batteries are inserted.
The fuse has burnt.
The AC adapter is not connected.
The plug of the AC adapter is not inserted
into the outlet, or there is no electricity in the
outlet.
The fuse has burnt.
The AC adapter’s fuse has burnt.
No batteries are inserted.
The batteries are fully charged or in a state
near full charge.
The temperature of batteries is low.
The temperature of batteries is high.
An error code is displayed.
Check the connection.
Check the connection.
Charge the batteries or use the AC adapter.
Insert batteries, or use the AC adapter.
Contact Brasseler USA.
Check the connection.
Check the connection.
Contact Brasseler USA.
Contact Brasseler USA.
Insert batteries.
No problem
If the temperature of batteries is less than
0°C (32°F), the batteries are not
rechargeable. Charge the batteries in a warm
room. (Be careful about moisture
condensation.)
It is normal that the batteries become a little
bit warm right after charging. If the batteries
are hot under normal operating conditions,
not right after charging, there may be an
abnormality. Contact Brasseler USA.
Refer to P.16 10. ERROR CODE.
The motor handpiece does
not rotate.
The motor handpiece does
not rotate. (The error code
“ ” is displayed.)
The motor handpiece cord is not connected.
The Foot Control (Option) is not connected.
There is a break in the motor handpiece or
the motor handpiece cord.
The contra angle head is clogged.
There is a short circuit inside the motor
handpiece or the motor handpiece cord.
Check the connection.
Check the connection.
Contact Brasseler USA.
Clean or replace the contra angle head.
Contact Brasseler USA.