background image

      Spray Nozzle

Spray Lubricant

Contra Angle Head

14

Autoclaving

1) Brush the debris off the surface of the contra angle head, and wipe it off with a cotton cloth moistened with isopropyl 

alcohol. Do not use metal brush.

2) Lubricate the head with the spray lubricant.

(Refer to the “P.14  7. (1) Lubricating Contra Angle Head.”)

3) Insert the head into an autoclave pouch and seal it.
4) Autoclavable up to max. 135°C (275ºF). 

ex.) Autoclave for 20 min. at 121°C (250ºF), or 15 min. at 132°C (270ºF).

5) Keep the contra angle head in the autoclave pouch to keep it clean until you use it.
* Sterilization at 121°C (250ºF) for more than 15 minutes is recommended by EN13060 or EN ISO17665-1.

• Sterilize the contra angle head only.
• For the sterilization method, we recommend the autoclave sterilization method.
• Sterilization is required first time you use and after each patient as noted below.

8. STERILIZATION

(1) Lubricating Contra Angle Head

• Lubricate the contra angle head only. 
• Supply spray after each use and/or before each calibration 

and autoclaving. 

1) Screw the spray nozzle onto the spray nozzle.
2) Insert the spray nozzle into the rear part of the contra angle 

head and lubricate the head for 2-3 seconds. (Fig. 12)

(2) Cleaning Motor Handpiece

When the motor handpiece becomes dirty, wipe it off with a cotton cloth moistened with isopropyl alcohol.

7. CLEANING

• Do not lubricate the motor section of the motor handpiece.
• To clean the motor handpiece, do not use any solvent such as benzine and, thinner.
• Before mounting the lubricated contra angle head to the motor handpiece, wipe off extra oil. Stand on its end or 

lean it in the proper position for gravity draining. Mount it after excess oil has been drained.

• Hold the contra angle head securely to prevent it from flying off by the pressure of the spray.
• Supply lubricant until it comes out of the handpiece head (for approx. 2 seconds).
• Do not use a spray can upside down. In such a case, only spray gas comes out and oil is not sprayed.

CAUTION

(5) Alarm Sound Volume Control

The alarm sound volume can be controlled at the three steps of “BIG VOLUME, SMALL VOLUME and LIMITED OFF. ”
1) Keep pressing both buttons of the PROGRAM key simultaneously.
2) The alarm mark on the LCD panel changes, and the sound volume changes.
3) Release the PROGRAM key when the sound volume reaches your required level and the required sound level is set up.

• The sound level is retained, even if the power is turned off.
• If P.12  6. (2) “Initialization of Program (Factory-set configuration)” is executed, the sound is set to “BIG VOLUME.”

NOTICE

Fig. 12

Содержание EndoDrive PLUS

Страница 1: ...Please Read this operation manual carefully before use and keep for future reference OPERATION MANUAL OM E0202E 001 Motor Handpiece System for Root Canal Enlargement ...

Страница 2: ... instructions care and maintenance Keep this Operation Manual for future reference Read these safety cautions thoroughly before use and operate the product properly These indicators are to allow you to use the product safely and prevent danger and harm to you and others These are classified by degree of danger damage and seriousness All indicators concern safety and should be followed Cautions for...

Страница 3: ...y result in personal injury due to its unexpected breaking or flying off during rotation Do not use or leave the product in a high temperature environment such as under strong direct sunlight in a car under a blazing sun by a fire or near a stove Check the product before use paying attention to looseness vibration noise and temperature heat generation If any abnormal condition is found even slight...

Страница 4: ...not used It is recommended to recharge the batteries just before use When the motor handpiece stops automatically by detecting a low battery voltage turning on the power again after a while may not detect the low voltage immediately This is not a failure but due to battery characteristics The voltage drop does not coincide with the remaining battery capacity Recharge rechargeable batteries after t...

Страница 5: ...13M DC7V 0 4A Ø20 L108mm 92g including motor handpiece cord Fig 1 2 3 8 10 9 1 12 4 5 6 7 11 Control Unit Operation Panel See Fig 2 Operation Panel and LCD Panel LCD Panel See Fig 2 Operation Panel and LCD Panel Error codes are displayed Power Cord Connector The AC adapter is connected Foot Control Connector The foot control option is connected 1 2 3 4 5 Motor Handpiece Connector The motor handpie...

Страница 6: ...on the rotation speed Operation Panel and LCD Panel Fig 2 LCD Panel This indicates that all displays are lighted TORQUE Key The torque limit values can be changed If the speed is to exceed the upper limit or fall below the lower limit the alarm sounds LCD The present torque limit value is displayed The unit is Ncm When it reaches the maximum value is displayed REVERSE Key REVERSE Key TORQUE Key TO...

Страница 7: ... available at 1 1 4 1 10 1 16 1 and 20 1 REVERSE Key The rotational direction of the file can be changed by this key It can also be changed while the file is in motion The alarm sounds during reverse rotation LCD The current rotational direction is displayed GEAR RATIO Key A gear ratio can be selected by this key If the correct gear ratio is selected according to the contra angle head in use an ad...

Страница 8: ...d regardless of the ON or OFF position of the power LCD The battery mark will always be displayed even if the unit is powered OFF The mark will be animated when the batteries are being charged or in the refresh mode Refer to P 13 6 3 Battery Refresh AC IN Lamp Green The lamp light will always be displayed when power is supplied by the AC adapter even if the unit is powered OFF AUTO REVERSE If the ...

Страница 9: ...nector with the mark of the plug down Fig 5 The Foot Control is optional and is not required to operate the EndoDrive PLUS 4 HOW TO CONNECT EACH PART To remove the plug hold the plug ring and pull it out NOTICE Bar Display LCD The bar display graph which shows the degree of load applied to the motor while the file is in motion Fig 3 Fig 5 Foot Control option Plug Foot Control option Plug Foot Cont...

Страница 10: ... especially prone to significantly shorter charging and operating times The temperature of the batteries is measured during charging Proper charging cannot be preformed if the charger is placed in an environment which is subject to sharp temperature change next to window subject to direct sunlight near air outlet or fan heater Place it in a place where temperature change is minimal Charging may no...

Страница 11: ...he GEAR RATIO key If the correct gear ratio is selected according to the contra angle head in use an adequate rotation speed and torque limit value can be set 4 Select the rotational direction by pressing the REVERSE key The key changes over to the forward rotation F and reverse rotation R each time it is pressed 5 Select the auto reverse mode by pressing the AUTO REVERSE key The key changes over ...

Страница 12: ...ou want it to rotate again re press the ON OFF key or re step on the foot control Optional 5 Operation 1 Start the motor handpiece to pressing the ON OFF button briefly Press the button again to stop the handpiece Alternate operation Keep pressing ON OFF button more than a second the motor handpiece starts while the button is pressed If you release the button it stops Intermittent operation If the...

Страница 13: ...for more than one second If the program number is changed by the PROGRAM key without pressing the MEMORIZE key the initially memorized preset values remain Cancel function NOTICE 6 Completion of Medical Treatment When the treatment is completed return the motor handpiece to the handpiece stand and hold down the POWER key for more than one second to turn off the power If 10 minutes pass without ope...

Страница 14: ...he time of charging At the same time the CHARGE lamp flashes slowly 5 Leave it as it is The batteries are discharged and recharged automatically 6 Hold down the POWER key for a while when you want to stop this process At this time the power does not turn on Lubricate the contra angle head before Calibration This function does not operate if AC adapter is not used Do not touch or apply a load to th...

Страница 15: ...or 2 3 seconds Fig 12 2 Cleaning Motor Handpiece When the motor handpiece becomes dirty wipe it off with a cotton cloth moistened with isopropyl alcohol 7 CLEANING Do not lubricate the motor section of the motor handpiece To clean the motor handpiece do not use any solvent such as benzine and thinner Before mounting the lubricated contra angle head to the motor handpiece wipe off extra oil Stand o...

Страница 16: ...13 3 Take out the old batteries and pull out the cord holding it at the connector Fig 14 4 Insert the connector of the cord of new batteries into the connector in accordance with the polarity indication label inside the battery chamber and place the batteries into the chamber with care so as not to catch the cord 9 CHANGING BATTERIES In case of difficult to insert the connector onto the battery th...

Страница 17: ...nction of the circuit may be suspected Contact Brasseler USA Malfunction of circuit The motor handpiece is locked at the time of the auto reverse mode Malfunction of circuit High load was continuously applied to the motor handpiece for a relatively long time Malfunction of circuit The motor handpiece is locked at the time of startup The voltage of batteries is too low The life of batteries or not ...

Страница 18: ...teries is high An error code is displayed Check the connection Check the connection Charge the batteries or use the AC adapter Insert batteries or use the AC adapter Contact Brasseler USA Check the connection Check the connection Contact Brasseler USA Contact Brasseler USA Insert batteries No problem If the temperature of batteries is less than 0 C 32 F the batteries are not rechargeable Charge th...

Страница 19: ...d on while pressing the ON OFF button There is a short circuit inside the ON OFF button The power is turned on while pressing the Foot Control Option There is a short circuit inside the Foot Control Option It is rotated by the ON OFF button It is rotated by the Foot Control Option Clean or replace the contra angle head Contact Brasseler USA Check the ON OFF button Contact Brasseler USA Check the F...

Страница 20: ...y electronic equipment The EndoDrive PLUS is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supply network that supplies buildings used for domestic purposes TUV Rhineland of North America is a Nationally Recognized Testing Laboratory NRTL in the United States and is accredited by the Standar...

Страница 21: ...lectromagnetic immunity The EndoDrive PLUS is intended for use in the electromagnetic environment specified below The customer or the user of the EndoDrive PLUS should assure that it is used in such an environment NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection...

Страница 22: ...prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the EndoDrive PLUS as recommended below according to the maximum output power of the communications equipment For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equ...

Страница 23: ... Specifications may be changed without notice ByYour Side in Dentistry One Brasseler Boulevard Savannah GA 31419 To order call 800 841 4522 or fax 888 610 1937 Visit our website www BrasselerUSA com ...

Отзывы: