21
Complies with
Maximum length
Cables and accessories
RF emissions, CISPR11,
Harmonic emissions,
Voltage fluctuations/ flicker emission
Electrostatic discharge (ESD)
Surge
Voltage dips, short interruptions and voltage variations on power supply input lines
Power frequency (50/60Hz) magnetic field
Conducted RF
Radiated RF
1.5 m
2.5 m
Plug side : 1.8m
Unit side : 1.8m
Motor Handpiece cord
Foot Controler cord
AC Adapter
Class B/ Group 1
IEC61000-3-2, Class A
IEC61000-3-3
IEC61000-4-2
IEC61000-4-5
IEC61000-4-11
IEC61000-4-8
IEC61000-4-6
IEC61000-4-3
150kHz to 80MHz
d =1.2 P
Recommended separation distances between portable and mobile RF communications equipment and the EndoDrive Plus.
The EndoDrive PLUS is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the
EndoDrive PLUS can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the EndoDrive PLUS as recommended below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Rated maximum output power of transmitter
W
800MHz to 2.5GHz
d =2.3 P
80MHz to 800MHz
d =1.2 P
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
m
0.01
0.1
1
10
100
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23