Inadequate battery
capacity
When the low battery voltage is below the alarm voltage, if the battery in
use is alkaline battery, replace the battery immediately. If the battery in
use is a lithium battery, connect the charger base and AC power supply to
recharge the lithium battery, and then use the battery to supply power as
needed.
D Manufacturer’s Declaration of the EUT
Guidance and manufacturer´s declaration – electromagnetic emission –
for all EQUIPMENT AND SYSTEMS
1
Guidance and manufacturer´s declaration – electromagnetic emission
2
The Palm Pulse Oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of Palm Pulse Oximeter should assure that it is used in such an
environment.
3
Emissions test
Compliance
Electromagnetic environment - guidance
4
RF emissions
CISPR 11
Group 1
The Palm Pulse Oximeter uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
5
RF emissions
CISPR 11
Class B
The Palm Pulse Oximeter is suitable for use in all establi
shments, including domestic establishments and those dir
ectly connected to the public low-voltage power supply ne
twork that supplies buildings used for domestic purposes.
6
Harmonic
emissions
IEC 61000-3-2
N/A
7
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
N/A