29
disposed of.
During the measurement temperature measurement function self-test
automatically once per hour. Self-Test for 2 seconds, will not affect the
normal operation of the oximeter.
7.3 On / Off Parameters Alarm
1. Select
【
Menu
】
→
【
Alarm
】
→
【
Alarm temperature alarm
】
2. Set the
【
Alarm Temp
】
as follows:
【
Open
】:
When this parameter alarm occurs, the oximeter will
alarm-related parameters;
【
Off
】:
oximeter without alarm-related parameters, alarm sounds, alarm
lights and alarms are turned off, the display Temp parameter shows “
”
icon.
7.4 Set the Alarm Limit
1. Select
【
Menu
】
→
【
Alarm
】
→
【
Alarm temperature alarm
】
2. Set Temp The
【
High Limit
】:
When the measured parameter value is above
the high alarm limit, an alarm will be triggered physiological parameters too
high.
3. Set Temp the
【
Lower limit
】:
When the measured value is below the low
alarm limit parameters, low physiological parameters will trigger an alarm.
When a parameter alarm occurs, This value will flash and corresponding
trigger an alarm lamp.
7.5 Setting The Temperature Unit
Select
【
Menu
】
→
【
Settings
】
→
【
TEMP unit
】,
options are "
℃
" and "
℉
"