7 Operation
7.1 Before use
The instrument is assembled, see section
1. Switch on the HF device and connect the HF cable to the HF device.
2. Set the required settings on the HF device.
3. Always perform a thorough visual inspection and function check before use (see section
7.2 Function test in the operating theatre
1. Check whether the jaws can be easily opened and closed using the handle.
2. Check the activation using the finger switch or foot switch.
The activation signal sounds when the finger switch or foot switch is activated.
WARNING
Risk of patient injury!
Use only compatible HF generators (see BOWA declaration of
compatibility).
Select the settings of the HF device to the mode required for the
operation.
Use only suitable devices and accessories wherever possible.
Use only intact and sterilised devices.
WARNING
Risk of patient injuries from burns or explosion!
Avoid contact with flammable gases and liquids, such as skin
cleansers, disinfectants and anaesthetic gases.
Avoid direct skin contact with HF cables.
Skin-to-skin contact (e.g. between the patient's arms and body)
should be avoided, e.g. by inserting dry gauze.