Pre-conditioning
BOWA-IFU-12578-ERGOACT-S0-EN-20210715
Instructions for Use ERGOact
19
9 Pre-conditioning
9.1 General information
The following values for the permitted number of pre-conditioning cycles are intended as
a guide. The actual number may vary depending on the demands placed on the devices.
Manual pre-conditioning methods are not recommended due to their significantly lower
effectiveness.
Follow the chemical manufacturers’ statements on concentrations and exposure times.
The medical device manufacturer cannot be held liable if other types of cleaning agents
and disinfectants are used.
9.2 Device-specific information
Insulated parts must not come into contact with hard, sharp or heavy products, as this
could damage the electrical insulation.
CAUTION
Risk of infection from water sprayers and steam during the manual
pre-cleaning process!
Wear a face mask and protective clothing.
Adequate ventilation is recommended.
NOTE
Abrasives and metal brushes cause damage to the jaws!
Never clean the laparoscopy instrument using abrasives or metal
brushes.
NOTE
Compressed air causes damage to the instrument parts!
Restrict compressed air pressure to 3 bar or less for drying the
instrument parts.