Operation
16
Instructions for Use ERGOact
BOWA-IFU-12578-ERGOACT-S0-EN-20210715
3. Place the tissue to be sealed between the jaws.
4. Close the jaws to grasp the tissue.
Never grasp too much tissue.
The tissue is grasped and pinched.
5. Using the finger switch or the foot switch, activate the HF current of the HF generator
for coagulation:
A continuous acoustic signal sounds while the HF current is activated.
6. Release the actuated switch.
The tissue is coagulated.
The tissue has been grasped by the jaws and is coagulated.
7.4 Removal
1. Close the jaws by closing the handle.
2. Withdraw the instrument from the trocar sleeve.
WARNING
Incomplete coagulation!
Never use the AUTOSTART function.
Do not switch on the HF current before the jaws are in contact with the
tissue to be coagulated.
Activating the instrument accidentally can cause injuries to the patient.
WARNING
Incomplete coagulation due to contaminated jaws!
Clean the jaws using a damp cloth or a sterile plastic brush on a regular
basis if tissue adheres to them.
WARNING
Risk of patient injury due to damaged or broken-off parts!
Always check the instrument before use. All parts must be present.