Operation
BOWA-IFU-12578-ERGOACT-S0-EN-20210715
Instructions for Use ERGOact
15
7.3 During the operation
Introducing the instrument
1. Close the handle in order to keep the jaws closed.
2. Insert the instrument into the trocar sleeve.
Grasping, clamping and sealing tissue
1. Position the branches at the operation site.
2. Turn the rotation wheel (with the jaws open) to adjust the angle of the jaws.
WARNING
Risk of patient injury due to tissue clamping, especially with limited
visibility! Prepare the tissue to be coagulated by freeing it as far as
possible to prevent it from becoming pinched!
Perform the operation only with adequate visibility.
WARNING
Risk of patient injuries from hot jaws and release of steam!
Instrument tips may still be hot immediately after HF power has been
switched off.
Maintain sufficient distance between instrument tips and sensitive
tissue structures, such as nerves, pancreas or intestines.
Make sure that no laparoscopy instruments under thermal load are
used to prepare the tissue.
Do not lay any laparoscopy instrument on the patient.
WARNING
Risk of patient injury due to accidental activation of the laparoscopy
instrument!
Never use the AUTOSTART function.
Do not switch on the HF current before the jaws are in contact with
the tissue to be coagulated.
Activating the instrument accidentally can cause injuries to the
patient.