Pre-conditioning
BOWA-IFU-12578-ERGOACT-S0-EN-20210715
Instructions for Use ERGOact
21
9.3 Inspection
These devices are subject to wear when used as intended, depending on the intensity of
use. This wear arises from the design and construction of the devices and is unavoidable.
Replace the device if it has externally visible defects or does not function as described in
this manual. Please advise the manufacturer or the manufacturer’s authorised
representative in such cases.
After cleaning and disinfecting the device, perform a visual inspection and function
check of the individual instrument parts.
Replace any damaged parts.
Handle
1. Check the actuation lever for ease of motion.
2. Check the insert for damage and corrosion.
Shaft tube
1. Check the insulation of the shaft tube for damage.
Re-conditioning step Description
Steam sterilisation
7
Steam
sterilisation
5
Sterilise the product while assembled or dismantled.
BOWA recommends steam sterilisation using the fractionated vacuum method
with a sterilisation time of 3–20 minutes and a temperature of 134–137°C in a
suitable sterilisation packaging.
BOWA has validated:
1 Soaking in cold tap water for 5 minutes
2 Ultrasonic bath for 5 minutes with Gigasept® InstruAF (Schülke & Mayr GmbH), 3% v/v
3 Neodisher® MediClean forte (Chemische Fabrik Dr. Weigert), 0.5% v/v
4 Thermal disinfection at 90°C for 5 minutes
5 Steam sterilisation: Half-cycle with 3 pre-vacuum phases at 132°C for 1.5 minutes, instrument
assembled, packaging: double in paper/foil sterilisation bag
CAUTION
Brittle or defective insulation can burn the patient!
Replace instrument components with damaged insulation.