INITIAL INSPECTION
When you first unpack your Bovie
®
IDS-400, inspect it visually:
• Look for any signs of damage.
• Verify that the shipping package contains all items listed on the packing list.
If the unit or any accessories are damaged, notify Bovie’s Customer Service immediately.
Do not use any damaged equipment.
INSTALLATION
Place the Bovie
®
IDS-400 on any flat surface with a tilt angle not more than 10˚. The unit relies on natural convection cooling. Do
not block its bottom or rear vents. Ensure that air flows freely on all sides of the unit.
WARNING:
Connect the power cord to a properly polarized and grounded power source with the frequency
and voltage characteristics that match those listed on the back of the unit.
FUNCTION CHECKS
Upon initial installation of the unit, perform the tests listed below. Refer to the figures in the previous chapter for the location of
connectors and controls.
WARNING:
At no time should you touch the active electrode or bipolar forceps. A burn could result.
Setting Up the Unit
1. Verify that the Power Switch is in the Off (O) position and that no accessories are connected to the unit.
2. Connect a hospital grade power cable to the AC power cable receptacle on the back of the unit, then to a properly grounded
wall outlet.
3. Connect a two-button monopolar pencil to the appropriate receptacle.
4. Do not connect a patient return electrode at this time.
5. Turn the unit on by switching the power switch to the On (|)position.
Checking the Return Electrode Alarm
1. Adjust the power settings for each mode (Cut, Coag, Bipolar) to one watt.
2. Press the Coag button of the pencil. Verify that an alarm sounds for three seconds and the patient return electrode sensing
alarm indicator light illuminates, indicating that no return electrode is connected to the unit.
3. Verify that adjusting the volume control on the back of the unit while the alarm is sounding does not change the alarm volume.
3-2
Bovie Medical Corporation
Содержание ids-400
Страница 1: ......
Страница 2: ...i User s Guide Bovie IDS 400 USER S GUIDE...
Страница 3: ...ii Bovie Medical Corporation...
Страница 15: ...1 8 Bovie Medical Corporation...
Страница 29: ...3 4 Bovie Medical Corporation...
Страница 45: ...7 4 Bovie Medical Corporation...
Страница 61: ...A 16 Bovie Medical Corporation...
Страница 64: ......