Setting Your Presets
Select the desired preset (0-9) by pressing the recall button.
Select the desired modes to be stored by pressing the mode membrane switches (Cut and Coag).
If presetting the Blend mode, select the desired level of hemostasis (Blend Bar 1-10) by pressing the
Blend amount control button.
Select the desired power (Cut, Coag, and Bipolar) to be stored by using the power output up and down membrane switches.
Once all of the settings are selected, depress and hold the Set button for three seconds. To indicate the settings have been
stored, the Preset Memory Number (0-9) will blink.
To recall a Preset, repeatedly press the Recall button to toggle through all of the presets.
NOTICES:
The IDS-400 incorporates 10 factory-set presets that are all set to zero and can be reset to your
preferred settings.
A small red dot blinking in the lower right corner of the Preset indicator display indicates that the
unit is not presently set to a user-defined preset.
Set and Recall buttons are disabled while the unit is activated.
Presets only store one Cut mode (Cut I or Cut II, or Blend) and power setting, one Blend level (if
applicable), one Coag mode (Pinpoint or Spray) and power setting, and Bipolar power setting.
When storing, only the information displayed in the display windows will be saved to the unit’s
memory.
Memory Feature (Last Selected Preset)
The Memory feature allows the unit to display the last selected power preset when the generator is turned on.
NOTICE:
To have a setting selection available at startup or to be one of the 10 user-defined presets, the
adjustment to the mode and/or power settings must be saved by pressing the Set button on the
Preset display panel.
Examples
Examples 1 through 4 explain how the Memory and temporary memory features work and what happens when the power and/or
mode is adjusted but not saved as one of the 10 Preset selections. Example 5 explains what happens when the power and/or mode is
adjusted and saved as a new Preset setting:
#1. The physician performs a surgical procedure using Preset 2. The Preset has been stored with the following mode and power:
• The mode is set to Cut I
• The power setting for Cut I is 30 watts
• The power setting for Pinpoint is 15 watts
• The power setting for Bipolar is 20 watts.
The procedure is completed and the unit is switched off.
The next time the unit is switched on, the number 2 Preset will be displayed and available when the unit is activated. The
number 2 Preset will be the same as the modes and settings indicated above.
#2. The physician performs a surgical procedure using Preset 2 (same as Example #1 Preset values).
He adjusts the power settings for each mode but does not store the new settings into the Preset.
The next time the unit is switched on, the number 2 Preset will be displayed and available when the unit is activated. The
number 2 Preset will be the same as the modes and settings indicated in Example #1.
4-6
Bovie Medical Corporation
Содержание ids-400
Страница 1: ......
Страница 2: ...i User s Guide Bovie IDS 400 USER S GUIDE...
Страница 3: ...ii Bovie Medical Corporation...
Страница 15: ...1 8 Bovie Medical Corporation...
Страница 29: ...3 4 Bovie Medical Corporation...
Страница 45: ...7 4 Bovie Medical Corporation...
Страница 61: ...A 16 Bovie Medical Corporation...
Страница 64: ......