OUTPUT CHARACTERISTICS
Maximum Output for Monopolar and Bipolar Modes
Power readouts agree with actual power into rated load to within 20% or 5 watts, whichever is greater.
EMC COMPLIANCE
Special precautions should be taken regarding the IDS-400. Medical Electrical Equipment needs special precautions
regarding EMC and needs to be installed and put into service according to the EMC information provided in this
manual.
Understand that only the Accessories supplied with or ordered from Bovie should be used with your device. The use
of accessories, transducers, and cables other than those specified, may result in increased Emissions or decreased
Immunity of the IDS-400. The IDS-400 and its accessories are not suitable for interconnection with other equipment.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment. The IDS-400 should
not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the IDS-400
should be observed to verify normal operation in the configuration in which it will be used.
A-5
User’s Guide • Bovie
®
IDS-400
Mode
Output Power
Output Frequency
Repetition Rate
Vpeak max
Crest Factor*
(Rated Load)
Cut I
400 W @ 500 Ω
490 kHz ± 4.9 kHz
N / A
1000V
1.6 ± 20%
Cut II
300 W @ 300 Ω
490 kHz ± 4.9 kHz
N / A
625V
1.6 ± 20%
Blend (Max)
200 W @ 300 Ω
490 kHz ± 4.9 kHz
30 kHz ± 5 kHz
2000V
3.5 ± 20%
Pinpoint
120 W @ 500 Ω
490 kHz ± 4.9 kHz
30 kHz ± 5 kHz
2400V
4.5 ± 20%
Spray
80 W @ 500 Ω
490 kHz ± 4.9 kHz
30 kHz ± 5 kHz
4000V
6.5 ± 20%
Bipolar
120 W @ 150 Ω
490 kHz ± 4.9 kHz
N/A
450V
1.6 ± 20%
Содержание ids-400
Страница 1: ......
Страница 2: ...i User s Guide Bovie IDS 400 USER S GUIDE...
Страница 3: ...ii Bovie Medical Corporation...
Страница 15: ...1 8 Bovie Medical Corporation...
Страница 29: ...3 4 Bovie Medical Corporation...
Страница 45: ...7 4 Bovie Medical Corporation...
Страница 61: ...A 16 Bovie Medical Corporation...
Страница 64: ......