PERFORMANCE CHARACTERISTICS
Input Power
Duty Cycle
Under maximum power settings and rated load conditions (Pure Cut, 400 watt @ 500 ohm load), the generator is suitable for
activation times of 10 seconds ON followed by 30 seconds OFF for one hour.
The internal temperature of the unit is continuously monitored. If the temperature rises above 85
0
C, the alarm will sound and
output power will be deactivated.
Dimensions and Weight
Operating Parameters
Transport and Storage
Generator should fit on all standard Carts for monopolar generators. The device should be stored and used in a room temperature of
approximately 77
0
F/25
0
C.
A-2
Bovie Medical Corporation
Ambient temperature range
-40˚ to +70˚ C
Relative humidity
10% to 100%,
including condensation
Atmospheric pressure
50kPa to 106kPa
Input Voltage
100-240 ~ VAC ± 10%
Mains line frequency range (nominal):
50 – 60 Hz
Power consumption:
560 VA
Fuses (two):
10 A (fast blow)
Width
31.1 cm (12.25 in.)
Depth
41.3 cm (16.25 in.)
Height
15.3 cm (6.00 in.)
Weight
< 8.75 kg (< 19 lbs)
Ambient temperature range
10° to 40° C
Relative humidity
30% to 75%, non-condensing
Atmospheric pressure
70kPa to 106kPa
Warm-up time
If transported or stored at temperatures outside the operating temperature range,
allow one hour for the generator to reach room temperature before use.
Содержание ids-400
Страница 1: ......
Страница 2: ...i User s Guide Bovie IDS 400 USER S GUIDE...
Страница 3: ...ii Bovie Medical Corporation...
Страница 15: ...1 8 Bovie Medical Corporation...
Страница 29: ...3 4 Bovie Medical Corporation...
Страница 45: ...7 4 Bovie Medical Corporation...
Страница 61: ...A 16 Bovie Medical Corporation...
Страница 64: ......