CAUTIONS:
Do not stack equipment on top of the generator or place the generator on top of electrical
equipment. These configurations are unstable and/or do not allow adequate cooling.
Provide as much distance as possible between the electrosurgical generator and other electronic
equipment (such as monitors). An activated electrosurgical generator may cause interference with
them.
Non-function of the generator may cause interruption of surgery. A backup generator should be
available for use.
Do not turn the activation tone down to an inaudible level. The activation tone alerts the surgical
team when an accessory is active.
When using a smoke evacuator in conjunction with the electrosurgical generator, place the smoke
evacuator a distance from the generator and set the generator volume control at a level that
ensures that the activation tones can be heard.
NOTICE:
If required by local codes, connect the generator to the hospital equalization connector with an
equipotential cable.
SETTING UP
1. Verify that the generator is Off by pressing the power switch Off (O).
2. Place the generator on a stable flat surface, such as a table, platform, or medical cart. Carts with conductive wheels are
recommended. For details, refer to the procedures for your institution or to local codes. Provide at least 10 to 15 cm (4 to 6 in.)
of space from the sides and top of the generator for cooling. Normally, the top, sides, and rear panel are warm when you use the
generator continuously for extended periods of time.
3. Plug the generator power cord into the AC Power Cable Receptacle on the rear panel.
4. Plug the generator power cord into a grounded receptacle.
5. Turn on the generator by pressing the power switch On (|). Verify the following:
• All visual indicators and displays on the front panel illuminate.
• Activation tones sound to verify that the speaker is working properly.
6. If the self-test is successful, a tone sounds. Verify the following:
• A Cut mode is selected; a Coag mode is selected.
• Each display shows a power setting. The unit automatically powers up to the last selected preset settings.
• The Patient Return Electrode Alarm Indicator illuminates red.
If the self-test is not successful, an alarm tone sounds. An error code will appear in the Bipolar display, in most cases, the generator
is disabled. Note the error code and refer to Section 6,
Troubleshooting
.
Once the self-test is successful, connect the accessories and set the generator controls. Refer to
Preparing for Monopolar Surgery
or
Preparing for Bipolar Surgery
later in this section.
4-3
User ’s Guide • Bovie
®
IDS-400
Содержание ids-400
Страница 1: ......
Страница 2: ...i User s Guide Bovie IDS 400 USER S GUIDE...
Страница 3: ...ii Bovie Medical Corporation...
Страница 15: ...1 8 Bovie Medical Corporation...
Страница 29: ...3 4 Bovie Medical Corporation...
Страница 45: ...7 4 Bovie Medical Corporation...
Страница 61: ...A 16 Bovie Medical Corporation...
Страница 64: ......