Precision™ Spinal Cord Stimulator System Clinician Manual
Clinician Manual
90607862-30 50 of 309
en
Specifications and Technical Data
Parameter
Range
Default
Areas (Channels)
4
—
Amplitude
0 – 20 mA
a
0 mA
Rate
2 – 1200 pps
b
40 pps
Width
0 – 1000 µsec
210 µsec
Cycle
0 – 90 min, OFF
OFF
Ramp ON
1 – 10 secs
3 secs
Contacts
1 – 16; +, -, OFF
1 – 16: OFF
a. The Precision System includes programmable coverage areas with each individ ual electrode contact limited to 12.7 mA.
A programming interlock is enforced to limit the coverage area output current to 20 mA or less. For example, a maximum
current output of 12.7 mA on a first electrode would limit the total summed current output on remaining electrodes to 7.3
mA within one coverage area.
b. Only one Area is available if the rate is 130 pps.
Materials
The Stimulator contains a radiopaque identification tag. The identification tag is visible using
standard x-ray procedures.
Case
Titanium
Header
Epoxy
Strain Relief
Silicone
Size/Volume
55 mm x 45 mm x 11 mm / 20.7 cm
3
(including header)
Содержание Precision Spinal Cord Stimulator System
Страница 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Страница 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Страница 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Страница 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Страница 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Страница 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...