Precision™ Spinal Cord Stimulator System Clinician Manual
Clinician Manual
90607862-30 26 of 309
en
Note:
• Ensure that the hex wrench is
fully seated in the setscrew before
tightening.
• The hex wrench is torque-limited
and cannot be overtightened.
4. Form an appropriately-sized pocket using
blunt dissection on either side of midline
for coiled excess lead and extension
connectors.
5. Place a small loop at the lead for slack.
If necessary, loosely tie a suture around
the lead-loop, but do not tighten onto the
lead.
CAUTION: Tightening sutures directly
on the lead can damage the lead.
6. Carefully remove excess slack by
gently pulling the extensions from the exit
wound.
7. For Permanent Trials, if desired, a small
suture may be used to close the exit
wound of the extension. Place and tape
a stress relief loop and dress the wound.
Proceed to the instructions for “Connect-
ing to the Trial Stimulator” page 32.
8. For Permanent Implant, close the midline
incision, and proceed to the instructions
Содержание Precision Spinal Cord Stimulator System
Страница 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Страница 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Страница 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Страница 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Страница 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Страница 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...