Sterilization, Handling, and Storage
Clinician Manual
en
Lead, Lead Extension, and
Splitter Handling and Storage
•
Avoid damaging the lead with sharp
instruments or excessive force during
surgery.
•
Do not sharply bend or kink the lead,
extension, or splitter.
•
Do not tie suture(s) directly to the lead,
extension, or splitter body; use the
provided suture sleeves.
•
For the Percutaneous Leads - avoid
forcing the lead into the epidural space
by carefully clearing a path using the lead
blank.
•
Avoid pulling an implanted lead taut;
provide a stress relief loop at the insertion
site to minimize tension on the lead.
•
Avoid handling the lead with sharp
instruments; use only rubber-tipped
forceps.
•
Take care when using sharp instruments
such as hemostats or scalpels to prevent
damaging the lead.
•
Wipe off any body fluids from the lead
connector end before connecting it to any
other component. Fluid contamination of
these connections could compromise the
integrity of the stimulation circuit.
•
Wipe off any body fluids from the lead
stylet before inserting or reinserting it into
the lead.
Store components between 0 °C and 45 °C
(32 °F and 113 °F) in an area where they are
not exposed to liquids or excessive moisture.
Temperatures outside of the stated range can
cause damage.
Содержание Precision Spinal Cord Stimulator System
Страница 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Страница 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Страница 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Страница 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Страница 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Страница 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...