Precision™ Spinal Cord Stimulator System Clinician Manual
Clinician Manual
90607862-30 16 of 309
en
lead stylet in the lead while the electrode
array is in the bevel of the insertion
needle increases the risk of lead and
tissue damage.
WARNING:
If the lead stylet is removed
and reinserted, do not use excessive
force when inserting the stylet into the
lead. The use of instruments, such as
forceps, to grasp the stylet during inser-
tion is not recommended as this could
result in applying excessive force and
could increase the risk of lead and tissue
damage.
7. Advance the lead to the appropriate ver-
tebral level under fluoroscopic guidance.
A sufficient length of lead (i.e., at least
10 cm, or approximately three vertebrae)
should reside in the epidural space to aid
in lead stabilization.
8. If use of a splitter is desired or you are
using the Infinion™ 16 lead, continue
“Lead Connection to Splitter” on page 21.
Otherwise, continue to “Connecting OR
Cable Assembly” on page 24.
Lead Connection to Splitter
1. Carefully withdraw the stylets from the
leads to be inserted into the splitter.
2. Wipe clean the proximal connector ends
of the leads.
3. Select the desired splitter model.
Note:
A Splitter 2x8 must be used when
implanting the Infinion™ 16 lead
4. Check that the lead connector end can
be easily inserted into the splitter without
obstruction. If obstruction is encountered,
loosen the set screws of the splitter by
using the hex wrench provided, turning
counterclockwise.
Note:
• The set screw should only be
loosened to an amount sufficient to
insert a lead.
• Do not excessively loosen the set
screw. This may cause the set screw
to dislodge rendering the splitter
unusable.
5. Insert proximal connector ends of desired
leads into splitter receptacles until they
are fully seated – each lead bottoms out
in receptacles and retention sleeves (long
ring) are under the setscrew blocks of the
splitter receptacles. Do not tighten set
screw at this time.
6. Continue to “Connecting the OR Cable
Assembly,” page 24, then proceed to
step 7 below.
7. Check connections with an impedance
measurement. If the impedance is
satisfactory, proceed to “Intraoperative
Stimulation Testing” on page 25 to confirm
proper lead location.
Содержание Precision Spinal Cord Stimulator System
Страница 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Страница 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Страница 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Страница 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Страница 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Страница 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...