Precision™ Spinal Cord Stimulator System Clinician Manual
Clinician Manual
90607862-30 44 of 309
en
3. Each option is discussed below.
Selection #1 – CP Mode
1. From the Clinician Options menu, high-
light “CP Mode” and press the
P
button
to select this menu item. This will prepare
the Remote Control for communication
with the Clinician Programmer (CP).
When in CP Mode, “CP Ready” will ap-
pear on the Remote Control screen.
2. Place the Remote Control and the IR
dongle in the plastic IR Holder so that
their communication ports are facing each
other.
3. Plug the dongle’s USB connector into the
appropriate connector on the CP.
4. Launch the Bionic Navigator software and
wait for the “Communication Established”
message.
Note:
• The Remote Control will remain CP
Ready for up to 15 minutes with no
activity.
• All buttons are active during CP
Ready and pressing any button
returns the Remote Control to the
Level screen.
• Stimulation may be turned on or off
in CP Ready state.
• Whenever the Remote Control is
re-activated from idle mode the
display will default to the Level
screen.
The Clinician Programmer can communicate
with either an External Trial Stimulator or
an IPG.
Содержание Precision Spinal Cord Stimulator System
Страница 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Страница 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Страница 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Страница 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Страница 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Страница 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...