Precision™ Spinal Cord Stimulator System Clinician Manual
Clinician Manual
90607862-30 32 of 309
en
Programming with the
Infinion™ 16 Lead
For detailed programming instructions, refer
to the appropriate Bionic NAVIGATOR™
Software Guide. When the Infinion™ 16
lead is used with the Splitter 2x8 and placed
properly into the Precision™ IPG ports
(laser-marked tail of Splitter 2x8 connecting
to IPG port “1-L”), the distal 8 contacts
of the Infinion™ 16 lead (contacts 1-8)
will correspond with contacts 1-8 on the
left side of the Bionic Navigator software
display. Likewise, the proximal 8 contacts
of the Infinion™ 16 lead (contacts 9-16) will
correspond with contact 9-16 on the right side
of the Bionic Navigator software display.
The diagram below depicts this wiring.
Using test stimulation, enlist patient feedback
to verify lead placement and pain coverage.
The user may program with the Infinion™
16 lead using all the functions available in
Bionic Navigator for the Linear™ family of
leads (E-Troll, Navigator, etc.). For complete
instructions on programming, see the Bionic
Navigator Software Guide.
Содержание Precision Spinal Cord Stimulator System
Страница 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Страница 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Страница 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Страница 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Страница 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Страница 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...