3
Focus
Changes distance between anodes and cathodes. A warning message will
appear if you try to steer when focus is “out of range”, and a blue vertical
line will appear in the focus window to indicate where it needs to be in order
to resume steering. The two focus lines (blue line, dark gray line) need to
overlap in order to steer current using the E-Troll and Navigate mode.
Understanding the Program Options
To view the Program Options:
1. Select the desired Program from the Program Carousel.
2.
Select
next to the Program Name.
Program Options Features
Copy to…
Copies the currently active Program to another slot in the Program Carousel.
Delete
Deletes all programming for the selected Program. After acknowledging, Area A will
be automatically created.
Revert to…
Gives the option of going back to the settings from the last programming visit (“Last
Clinician Settings”) or to the settings prior to making any changes in the current
programming visit (“Walk-in settings”).
Battery
Estimate
Displays an estimate of how often a patient with a rechargeable battery would have
to charge if the currently active Program were used.
Displays an Energy Use Index that corresponds to the estimated energy usage of a
non-rechargeable battery if the currently active Program were used. For additional
information, refer to the
Precision Novi Information for Prescribers Directions for Use.
Options
Opens up a new window where you can specify Stimulator, Program, and Area
options, such as Cycling, Maximum Amplitude, Ramp-Up time, and Remote Control
Locks.