Bionic Navigator™ 3D Programming Manual
Bionic Navigator™ 3D Programming Manual
90930923-31 REV A 44 of 309
en
5
Advanced
Options
Tab
When expanded, displays Resolution
and Focus controls. Refer to “Selecting
Advanced Options”.
Area
Tab (E-Troll™ and
Navigate Modes only)
6
Polarity and
Percentage Control
Use to change polarity (cathode or anode)
and percentage after selecting an individual
contact or the case.
Area
Tab (Manual Mode
only)
7
Equalize
Control
Use to equally distribute all cathodic or
anodic current among active contacts in the
selected Area
Area
Tab (Manual Mode
only)
Full View Tab
Expanding the Full View Tab will display the Lead Group panel, showing configured leads in each Lead
Group. This will help visualize what contacts are being used as you are programming.
Figure 25:
3
1
2
Full View Panel