Configuring the CP and Stimulator
Bionic Navigator™ 3D Programming Manual
90930923-31 REV A 29 of 309
en
The components on the LeadSync pop-up window are described in the table below.
Component
Description
1
Lead Group
drop-down list Allows you to select a specific Lead Group to review.
2
Cancel
button
Discards the LeadSync results for the Lead Group shown.
3
Rescan
button
Re - runs LeadSync for all Lead Groups.
4
Apply
button
Allows you to accept the LeadSync results for the Lead Group
shown, adjusting the relative alignment of the leads accordingly.
5
Close
button
Closes the LeadSync pop-up window to return to the Leads
Screen.
The image displayed on the LeadSync pop-up window may display a different relative lead orientation
than what was configured in the Leads Placement panel. Between each pair of leads is a symbol
indicating the correlation of the LeadSync measurement to the displayed lead offset, as the LeadSync
measurement is dependent on the geometry of the actual lead placement. There are three symbols that
describe LeadSync measurement results:
Correlation of LeadSync Measurement to the displayed lead offset is high. See Figure 18.
Correlation of LeadSync Measurement to the displayed lead offset is low. See Figure 19.
LeadSync is unable to provide meaningful result and will display the lead configuration that
was manually specified by user. See Figure 20.