2
1
(1) Preassembled tool holder
(2) Knurled screw
5. Position the mount for the line laser
(1)
with the
holes over the tube connectors on the guide tubes.
1
22
3
4
5
4 Nm
(1) Line laser mount
(2) Tube connectors on guide tubes
(3) Countersunk screw
(4) Washer
(5) Hex nut
6. Use the countersunk screws
(3),
washers
(4)
and
hex nuts
(5)
to secure the mount for the line laser on
the tube connectors at 4 Nm.
7. Slide the prismatic reflector
(1)
onto the tool holder.
4712008-04_shd
1
(1) Prismatic reflector
8. Secure the plastic screws
(1)
on the tool holder.
The plastic screws are intended as a stop so the
prismatic reflector cannot slide off the tool hold‐
er by accident. 3 plastic screws are included in
the scope of delivery. The third plastic screw is
intended as a spare.
4712008-05_shd
11
(1) Plastic screws
---Separator---
3.2
Checking the accuracy of the inclina‐
tion measurement (before each cali‐
bration)
Before each calibration, check the accuracy of the in‐
clination measurement. Use reverse measurement. To
do this, place the measuring instrument on a table
and measure the inclination. Turn the measuring in‐
strument by 180°, and measure the inclination again.
The maximum permissible difference between the
numbers indicated is
0.3°
. In case of a greater devia‐
tion, recalibrate the measuring instrument. To do this,
select CAL in the settings.
Follow the instructions on
the display.
We recommend a check of the accuracy
and, if necessary, a calibration of the measuring in‐
strument after severe temperature changes and im‐
pacts. After a change in temperature, give the measur‐
ing instrument some time to return to a normal tem‐
perature before performing the calibration.
---