       GasFinderFCr Operation Manual — Data I/O 

        July 17th 2017

Serial Communications

The 

GasFinderFCr

 will transmit serial data at 300 / 1200 / 2400 / 4800 / 9600/ 19200 / 38400 

/ 57600 / 115200 / 230400 baud, no parity, 8 bits, one stop bit, (i.e., 9600,N,8,1) after each 
scan. Two handshaking options are available:

No Handshaking — Connect pin 4 (PCDTR) to pin 12 (+5Vdc) at the 

• 

GasFinderFCr

 end 

of the data cable. Data will be sent automatically after every scan. This is the standard 
confi guration of the cable as supplied by Boreal Laser Inc.

DTR Handshaking — The control computer raises the voltage to +5 Vdc on pin 4 (PCDTR) 

• 

in order to request a data string. Data are transmitted as an ASCII string and will be sent only 
when requested by the host computer.

Data String Specifi cation:

$GFDTA

  

data_header

gas_concentration

confi dence_factor,R2  

distance

light_level

date_time

ser_number

status_code

check_sum 

header designating the type of data 

String

in ppmm (parts per million meter)     

Number

between 0 and 99     

 

 

Integer

user-set distance to retro in metres  

Number

returning light level    

 

 

Integer

current date and time               YYYY/MM/DD    hh:mm:ss
identifi cation number of the system 

String

code giving system operating status 

Integer

checks the data stream for errors  

String

Field Description 

   

 

Range      

Max. Width (Chars)   

Restrictions

data_header         

 

 

Any 

 

 

6  

gas_concentration         

0 –  99999999 

 

 

 

 

 

confi dence_factor       

      0 – 99 

 

 

2

distance              

       

    1 – 9999   

 

4

light_value 

       

      

   1 – 16384   

 

5

date_time        

 

Year/Month/Day  

 

19

 

 

       

 

Hour/Minute/Sec 

 

serial_number             

 

Any 

 

 

10

status_code         

       

     1 – FFFF   

 

4

check_sum         

                  0 – FF   

 

2

These data strings are comma-delimited (,)  and an asterisk (*) signifi es end of string.  
Each string is terminated by a carriage return and a line feed.
The status codes are described in detail in Appendix B of this manual. 

$GFDTA,  27.7, 98, 246.3, 5527,  2006/03/22 13:30:28, HFFC-1015,  1, *56

data_header

gas_concentration

confi dence_factor R2

distance

light_level

date_time

ser_number

status_code

check_sum

Содержание GasFinder FC

Страница 1: ...c GasFinderFCr Operation Manual July 17th 2017 GasFinderFCr Part No NDC 200005 D Menu ver 1 10 c 12846 146 Street NW Edmonton Alberta CANADA T5L 2H7 Phone 1 780 488 5173 Fax 1 780 488 0780 www boreal...

Страница 2: ...00010 C GasFinderPT NDC 200021 B GasFinder2 for Multiple Path Monitoring NDC 200025 A User Manuals GasViewOP NDC 200022 B GasViewMC NDC 200013 B GasMap NDC 200007 A Accessory Manuals 4 20mA Module NDC...

Страница 3: ...ent Check 2 2 Section 3 Installation 3 1 GasFinderFCr 3 1 Retroreflectors 3 1 Light Values 3 1 Condensation 3 3 Transceivers Remote Heads 3 4 OP1 Remote Head 3 4 OP3 Remote Head 3 5 Optical Alignment...

Страница 4: ...and Abbreviations 8 4 Section 9 Appendices 9 1 Appendix A Menu Structure 9 1 Appendix B Status Codes 9 3 Appendix C Explanation of ppmm 9 4 Appendix D Explanation of R2 9 5 Appendix E Conversion from...

Страница 5: ...ce Please install in accordance with these instructions WARNING TO REDUCE THE RISK OF ELECTRIC SHOCK Do to expose this equipment to rain or moisture Do not use the unit near water and do not immerse i...

Страница 6: ...real Laser Inc GasFinderFCr Operation Manual July 17th 2017 V1 Duct Probe GasFinderFCr mounted in cabinet 2 OP3 Remote Heads in stainless steel enclosures 2 OP1 Remote Heads OP1 Remote Head OP3 Remote...

Страница 7: ...derMC GasFinderAB and GasFinderPT laser gas detector products The warranty covers the GasFinder laser analyzers plus Boreal Laser manufactured items supplied with the GasFinder laser analyzers such as...

Страница 8: ...jects Extended warranty on GasFinder systems beyond the standard 12 month period PROCEDURES FOR RETURNING A UNIT TO BOREAL LASER Contact Boreal Laser Inc at support boreal laser com or 1 780 488 5173...

Страница 9: ...the GasFinderFCr is sent through a fibreoptic cable to the remote head or duct probe where it is emitted and propagates through the atmosphere to the retroreflector It then returns to the cell where i...

Страница 10: ...ad on the front panel or with a computer using GasViewOP These values are described in the Menu Structure which contains three subsections Main Menu contains items such as path lengths units memory et...

Страница 11: ...Belden 9880 Ethernet Cable Patch cord Belden 89907 Accessories OP1 Open Path Transceiver Remote Head Weight 3 0kg 6 6 lb Dimensions L x W x H 315 x 106 x 180mm 12 4 x 4 2 x 7 1ins Power Requirements...

Страница 12: ...ignition hazard presented by lasers of this type There are no user serviceable components contained in apparatus all units must be returned to factory for repair Caution The infrared laser output from...

Страница 13: ...and limiting paths to 500m for OP3 Remote Heads and be in areas that are safely accessible for installation and maintenance Limit cable runs to 750m or an appropriate distance to pull the cable Dusty...

Страница 14: ...ent wrench Small vise grips Phillips Robertson and Flat Head screwdrivers Allen Keys imperial and metric Coaxial cable termination tool Amphenol crimper ARF1118 with CTL 3 jaws 0 100D and 0 429D Wire...

Страница 15: ...signal return provided this does not degrade the R2 value Where retro reflectors are used on longer paths these may be mounted in an enclosure on a tripod or the enclosure placed in any convenient loc...

Страница 16: ...etros are probably the maximum number that can be economically used as the cost becomes prohibitive as the number of retros increases At path lengths approaching 750m maintaining the correct alignment...

Страница 17: ...This also prevents rain from striking the window Use a desiccant pack inside the enclosure This will only prevent condensation on the inside Place a heater in the enclosure The heater can be 12W to 50...

Страница 18: ...tate the unit so that the head is aligned with the reflector Tighten the 12mm bolt Remove the 8 screws from the rear plate Connect a visible laser to the fibreoptic laser connector at the rear of the...

Страница 19: ...especially before the shocks have been connected Remove the enclosure from the packing case open the cover and remove the base and the nose from inside There is a tubular spring loaded support laying...

Страница 20: ...X Y mount to the stainless steel bottom plate Bolt the base mount to a structurally stable base see page 7 5 and bolt the stainless steel enclosure with the Remote Head to the base mount Use the base...

Страница 21: ...than 2000 is an indication that the laser signal may have to be attenuated If the retroreflector is not visible in the alignment scope or difficulty is experienced getting a reading on the RLM connec...

Страница 22: ...g the eye laterally will cause the cross hairs to move across the Retro Remove the caps covering the adjustment screws Turn the zoom on the scope to a maximum Focus the eye piece so the retros are sha...

Страница 23: ...ted to prevent condensation forming on the window or the Retro Optical Alignment Remove the opposite port cover plate Attach a visible laser to the optical connector in the Duct Probe and adjust the a...

Страница 24: ...Kit contains a Remote Light Meter RLM Power Meter Visible Laser Infra Red IR card cables and cleaning tape Fibreoptic Connections The fibreoptic couplings are FC APC and are used to connect two FC AP...

Страница 25: ...cuts should be sharp and square Caution Do not nick the braid dielectric or centre conductor Do not comb out the braid 2 Insert the pin over the centre conductor so that it butts up against the dielec...

Страница 26: ...are usually spread spectrum type and can transmit data over distances limited by line of sight Cable Connections Power The GasFinderFCr will accept a power source at 100 250V 50 60 440Hz or 120 300V...

Страница 27: ...input from GasFinderFCr 4 PCDTR Data Terminal Ready Serial input from GasFinderFCr 5 GRD Signal Ground Digital Analogue 6 Ramp System Timing Clock Analogue output from GasFinderFCr 7 I sink 4 20mA loo...

Страница 28: ...time ser_number status_code check_sum header designating the type of data String in ppmm parts per million meter Number between 0 and 99 Integer user set distance to retro in metres Number returning l...

Страница 29: ...ing is taken from the GasFinderFCr internal clock 4 20mA Option Note that the 4 20mA output is taken from the pins in the DB15 output connector The 4 20mA output is configured using the GasFinderFCr k...

Страница 30: ...300m They both require a small amount of power to operate These are two small units one connected to the GasFinderFCr and the other to the computer with a short data cable They are both linked togeth...

Страница 31: ...cond baud rate to the correct value typically 9600 Also select 8 data bits no parity and 1 stop bit Flow control should be set to none Click OK If the computer is correctly connected to the GasFinderF...

Страница 32: ...sually 0 ppmm 4mA and 20mA varies proportionately to the gas concentration 20mA is equivalent to the maximum ppmm or greater If the Distributed Control System DCS is capable of a full 0 20mA resolutio...

Страница 33: ...is in the normal operating mode the system will enter the parameter display adjustment menus The arrow keys can be used to move between the various menu items When the operator is in the menu system p...

Страница 34: ...he system can operate well with a light value between 2 000 and 12 000 ppm HF R2 Calibrating Light Value Dist 6907 105 5 Concentration units ppm Main Menus Serial HFFC 1012 Version 1 10 c ppm HF R2 1...

Страница 35: ...stalled This is not an adjustable item 2 Concentration Units This item displays the units chosen to display the concentration in the main display The units can be adjusted to display ppmm parts per mi...

Страница 36: ...Boreal Laser Inc before changing this value from the factory default value 10b Min acceptable R2 This item is the same as item 6 but is for Gas B 11 Serial Output Menus This contains a submenu with tw...

Страница 37: ...rds when the memory is full Thus only the latest data is recorded Fill Alternatively if the fill mode is used the data will fill the buffer and once full no new data will be captured Downloading The b...

Страница 38: ...d the resulting readings are called sweeps A series of sweeps are averaged to form a scan When the data for the sample scan reference cell scan and associated readings are incorporated into the calcul...

Страница 39: ...ry Voltage This provides a convenient way of checking the voltage of the 12V power supply If the voltage falls below 11 5V a flashing B will appear in the top right corner of the display If the voltag...

Страница 40: ...h 2017 30 Service Menus This Submenu is only available to qualified personnel and requires a password to enter 31 Password Protected Menus These menus are accessible only for Boreal Laser personnel an...

Страница 41: ...on a clear optical path to obtain accurate readings of gas concentration Dust settling on the windows must be cleaned off whenever it degrades the instrument s performance The average value for the l...

Страница 42: ...n cleaning them The recommended procedure for cleaning is as follows Blow off any dust using canned compressed air A Note that compressed shop air may contain oil from the compressor and should not be...

Страница 43: ...once the obstruction has been removed e g fog lifts rain stops or vehicle equipment is moved No data transmission from GasFinderFCr to the computer Check that the correct baud rate is selected usually...

Страница 44: ...k on If the display still shows Cal Err collect an Arrays Transfer File page 7 4 and send it to Boreal Laser together with 30min of data from the gvl text file This will allow Boreal Laser to diagnose...

Страница 45: ...ding an array transfer file to an external computer The GasViewOP software can be used to record the file To download the arrays Connect the GasFinderFCr to a computer and run GasViewOP Click on Port...

Страница 46: ...centreline before the beam will move completely off the retro array This movement is referred to as translational In practice significant translational movement is unlikely to occur as the post is us...

Страница 47: ...o comply with safety regulations the height of a scaffold cannot exceed three times the width of the narrowest base dimension This gives a very stable platform however the alignment will often move sl...

Страница 48: ...the metal structure carrying the laser mount which twists with respect to the longitudinal axis of the original laser path This causes the laser beam to move away from the reflectors and results in a...

Страница 49: ...the light level drops Note where the cross hairs are in the scope relative to the reflector at these positions see figure 4 and position the head so that the cross hairs are in the centre of the trave...

Страница 50: ...7 8 Boreal Laser Inc GasFinderFCr Operation Manual Troubleshooting July 17th 2017...

Страница 51: ...principle except that the tube mirrors and crystal are all combined in a small package using semiconductor technology The diode laser is composed of a pn junction as in the light emitting diode LED an...

Страница 52: ...c absorption wavelength for this target gas See figure 2 The laser operating wavelength is kept stable by housing it in a temperature controlled box A fluctuating saw tooth current is used to drive th...

Страница 53: ...rear behind the glass See figure 4 A retroreflector is different in that the reflected light returns in the same direction as the incident light The beam is reflected 180 degrees A retroreflector is l...

Страница 54: ...ntroller A small device using a microcomputer microcomputer A small computer the size of a postage stamp micron 0 000 001 part of a meter modulated Like a wave molecule A collection of atoms bonded to...

Страница 55: ...ed A B B only Battery Voltage Gas B Detection Limits enter Adj Gas Type 12 35 A B escape Diabled A B B only esc esc ent esc esc esc esc esc ent esc esc esc ent esc esc esc Internal Temperature Min acc...

Страница 56: ...9 2 Boreal Laser Inc GasFinderFCr Operation Manual Appendices July 17th 2017...

Страница 57: ...1 1 C 1 1 0 0 D 1 1 0 1 E 1 1 1 0 F 1 1 1 1 n11 n10 n9 n8 Hex n15 n14 n13 n12 0 0 0 0 0 1 0 0 0 1 2 0 0 1 0 3 0 0 1 1 4 0 1 0 0 5 0 1 0 1 6 0 1 1 0 7 0 1 1 1 8 1 0 0 0 9 1 0 0 1 A 1 0 1 0 B 1 0 1 1 C...

Страница 58: ...a specific gas such as HF in the path of the laser beam between the transmitter unit and a reflector This is a total path measurement The units are parts per million metres or ppmm A uniform backgrou...

Страница 59: ...ed mathematical procedure to compare curves or numeric arrays is the Linear Least Squares Regression analysis This analysis results in a measure of the similarity R2 between the waveform of the sample...

Страница 60: ...he signal to noise ratio will increase The general shape of the plot is the same for all gases however the x axis values will depend on the sensitivity of the instrument to the gas species being obser...

Страница 61: ...m HF 0 8927 mg m3 at STP STP refers to Standard Atmospheric Pressure P 101 325kPa Standard temperature K 273 15K If the measurement is not taken at STP then an aditional correction is required to adju...

Страница 62: ...t item Log to Memory When this is Enabled the internal logging option is activated Log Memory Records This item lists the number of records currently in the memory The storage capacity is 131 072 line...

Страница 63: ...ans that the time for one sample is 1 1 second 33 30 50 samples will usually ensure that one reference cell check is included in the calculation The time for a one minute average will be 55 seconds 60...

Страница 64: ...of gas which is traceable to NIST standards The calibration data are stored in the instrument s software as a standard reference waveform During operation the GasFinderFCr s internal reference cell is...

Страница 65: ...Gas Shift Enable Enable Disable Centre Line Menu Attn Light Compensation output lin menu Min Attn Index Serial Password Protected menu Checked by Light Value Distance Ref Light Ref Cell ppmm Date Data...

Страница 66: ...ount Large Boreal X Y Mounta Stainless Steel Enclosure Enclosure for OP3 Remote Head Cables Data Cable Data Cable DB15M DB9F Data Cable DB15M DB9F with 4 20mA pair Fibreoptic Cable Preterminated with...

Страница 67: ...he signal by different amounts depending on the length of the cable The value of the compensation is specific to each channel and is entered at the time of commissioning by using the appropriate secti...

Страница 68: ...racters to keep the interface lean All commands are terminated by carriage return Enter Commands are responded to using the GasFinderFCr checksummed output line convention GF CER COK n XX where XX is...

Страница 69: ...0512FC A 2 GasFinderFCr Dimensions JB0125MC A 3 OP1 Head Overall Dimensions BOR 02110 J 4 OP3 Remote Head Dimensions JB 0930 RH B 5 V1 Duct Probe General DP D V1 001 6 V1 Duct Probe Dimensions DP D V1...

Страница 70: ...10 2 Boreal Laser Inc GasFinderFCr Operation Manual Drawings July 17th 2017 This page intentionally left blank...

Страница 71: ......

Страница 72: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...

Страница 73: ...BOREAL LASER INC MC Instrument Overall Dimensions JB0125MC A 1 of 1 NTS A Dim mm...

Страница 74: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...

Страница 75: ......

Страница 76: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...

Страница 77: ......

Страница 78: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...

Страница 79: ......

Страница 80: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...

Страница 81: ......

Страница 82: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...

Страница 83: ......

Страница 84: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...

Страница 85: ......

Страница 86: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...

Страница 87: ......

Страница 88: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...

Отзывы: