9.9
Boreal Laser Inc.
GasFinderFCr Operation Manual — Appendices July 17th 2017
Example 1:
If data are to be recorded every sample, ‘
Log to Memory’
should be
Enabled
, set the
‘Serial
Dump per Sample Sweep’
to
‘Serial Dump per Sample Sweep’
, and set the ‘
Average
Window’
to
1
.
Example 2:
To record data with a one-minute average, ‘
Log to Memory’
should be
Enabled
, set the
‘
Average Window’
to about
55
, and set the ‘
Serial Dump per Sample Sweep
’ to ‘
Serial
Dump per Average Window’
. In this confi guration, the recording time is nearly 3 months.
To calculate the correct fi gure for the sample:
Check the date/time entry in the ViewData window of
•
GasviewOP
. Note the time taken
for 50 samples to occur, say 55 seconds. This means that the time for one sample is 1.1
second(33/30). 50 samples will usually ensure that one reference cell check is included
in the calculation.
The time for a one minute average will be 55 seconds(60/1.1).
•
Note:
To ensure that the logging proceeds correctly, three menu items should be
checked:
‘Log to Memory’
(should be enabled),
‘Serial Dump per Sample/
Average Sweep’
(set to Average) and ‘
Average Window’ (
should be calculated
and entered).
When the
GasFinderFCr
is in
Menu Mode
, the data output is set automatically to once per
sample. This can appear confusing if the
Average Interval
is set to a value other than one.
However, when the
Menu Mode
is exited, data will be stored at the selected rate.
Care should be taken if the
GasViewOP
program is used at the same time as the
Log to
Memory
is enabled in an averaging mode. Due to possible re-averaging by the
GasViewOP
program, false values may be generated.
Содержание GasFinder FC
Страница 50: ...7 8 Boreal Laser Inc GasFinderFCr Operation Manual Troubleshooting July 17th 2017...
Страница 56: ...9 2 Boreal Laser Inc GasFinderFCr Operation Manual Appendices July 17th 2017...
Страница 70: ...10 2 Boreal Laser Inc GasFinderFCr Operation Manual Drawings July 17th 2017 This page intentionally left blank...
Страница 71: ......
Страница 72: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...
Страница 73: ...BOREAL LASER INC MC Instrument Overall Dimensions JB0125MC A 1 of 1 NTS A Dim mm...
Страница 74: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...
Страница 75: ......
Страница 76: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...
Страница 77: ......
Страница 78: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...
Страница 79: ......
Страница 80: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...
Страница 81: ......
Страница 82: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...
Страница 83: ......
Страница 84: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...
Страница 85: ......
Страница 86: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...
Страница 87: ......
Страница 88: ...Boreal Laser Inc GasFinderFCr Operation Manual Drawings January 2011 This page intentionally left blank...