manualshive.com logo in svg
background image

087-0065, Rev D/February 2019

WARNING: ALL INDIVIDUALS WHO HAVE OR WILL HAVE RESPONSIBILITY FOR

USING, MAINTAINING, OR SERVICING THIS PRODUCT MUST READ

THIS ENTIRE MANUAL CAREFULLY. FAILURE TO USE THIS EQUIPMENT

PROPERLY COULD RESULT IN SERIOUS INJURY OR DEATH.

3M™ PROTÉGÉ SG

PORTABLE, REUSABLE, SINGLE-GAS MONITOR

OPERATION AND MAINTENANCE MANUAL

087-0065, Rev D

Содержание PROTEGE SG

Страница 1: ...ESPONSIBILITY FOR USING MAINTAINING OR SERVICING THIS PRODUCT MUST READ THIS ENTIRE MANUAL CAREFULLY FAILURE TO USE THIS EQUIPMENT PROPERLY COULD RESULT IN SERIOUS INJURY OR DEATH 3M PROTÉGÉ SG PORTABLE REUSABLE SINGLE GAS MONITOR OPERATION AND MAINTENANCE MANUAL 087 0065 Rev D ...

Страница 2: ...019 Reformatted entire manual added new logo added INMETRO certifi cation updated flow charts in Section 2 updated appendices Document Title Document Number Purpose Protégé SG User Guide 087 0065 Provides information on operation and mainte nance on the Protégé SG monitor Masterdock II User Guide 087 0060 Provides information on operation and mainte nance on the Masterdock II station Protégé Gas D...

Страница 3: ...e 3M may make improvements or changes in the product s described in this documentation at any time While every effort has been made to ensure accuracy in this guide no responsibility can be accepted for errors or omissions This publication is not intended to form the basis of a contract and the com pany reserves the right to amend the design content and specifications of the detector without notic...

Страница 4: ...087 0065 Rev D February 2019 3M PROTÉGÉ SG PORTABLE REUSABLE SINGLE GAS MONITOR LEGAL STATEMENT iv This page is intentionally left blank ...

Страница 5: ...5 Span Calibration 2 4 2 6 Bump Test 2 5 3 INTRODUCTION 3 1 3 1 Monitor Overview 3 1 4 CONFIGURATION AND SETUP 4 1 4 1 Installing the PGD Application Software 4 1 4 2 Installing the USB Driver 4 1 4 3 Connecting the Device to a PC 4 4 4 4 Default Settings 4 6 4 5 Monitor Setup on PGD Application Software 4 7 4 6 Sensor Information 4 8 4 7 Monitor Configuration 4 8 5 OPERATION 5 1 5 1 Operating the...

Страница 6: ...t Section Title Page 6 2 Maintenance 6 6 6 2 1 REPLACING SENSORS 6 6 6 2 2 REPLACING THE LITHIUM BATTERY 6 8 6 2 3 REPLACING THE SENSOR FILTER 6 10 6 3 Error Codes 6 11 A SPECIFICATIONS A 1 B GAS INTERFERENCES B 1 C PARTS C 1 C 1 Applicable Parts List C 1 C 2 Sensor List C 2 D TECHNICAL SUPPORT D 1 ...

Страница 7: ... 4 2 4 3 USB Driver Installation 3 of 4 4 3 4 4 USB Driver Installation 4 of 4 4 3 4 5 Connecting the IR Connect to a PC 4 4 4 6 IRDA Indicator 4 4 4 7 Connected Monitors 4 5 4 8 PGD Application Software Dashboard 4 7 4 9 Sensor Information Window 4 8 4 10 Monitor Information Window 4 8 5 1 Monitor LCD Indicators 5 1 6 1 ZERO Calibration Screen 6 2 6 2 SPAN Calibration Calibration Adapter Attached...

Страница 8: ...itor 3 2 3 2 Monitor LCD Indicators 3 3 4 1 Installation Checklist 4 1 4 2 Factory Default Alarm Setpoints 4 6 4 3 Firmware Default Settings 4 6 5 1 Monitor LCD Indicators 5 2 5 2 Operation Buttons 5 2 5 3 Self Test Sequence 5 4 5 4 Monitor Alerts and Alarms Descriptions 5 6 6 1 Recommended Calibration and Test Matrix 6 1 6 2 Acceptable Battery Information 6 8 6 3 Error Codes 6 11 A 1 Monitor Spec...

Страница 9: ...nd Approvals for Monitor indi cate the monitor has been tested and complies with the following WARNING READ UNDERSTAND AND FOLLOW THE ENTIRE CONTENT OF THIS GUIDE PRIOR TO USE FAILURE TO DO SO MAY RESULT IN SERIOUS INJURY OR DEATH WARNING THIS ICON AND TEXT INDICATE A POTENTIALLY HAZARDOUS SITUATION WHICH IF NOT AVOIDED COULD RESULT IN DEATH OR INJURY CAUTION This icon and text indicate an action ...

Страница 10: ...IECEx certification standards and dates IEC 60079 0 2011 and IEC 60079 11 2011 Ex ia IIC T4 Ga IECEx certification standards and dates ABNT NBR IEC 60079 0 2013 and ABNT NBR IEC 60079 11 2013 Ex ia IIC T4 Ga DNV 17 0060 X Note The monitors covered in this manual have not been evaluated for enriched oxygen atmo sphere 21 Table 1 1 Certifications and Approvals Mark ...

Страница 11: ...odifications not expressly approved by the party responsible for compliance could void user s authority to operate the equipment This portable transmitter with its antenna complies with FCC s RF exposure limits for general pop ulation uncontrolled exposure The Protégé Plus Integrated Transmitter has been assigned FCC ID T5E Compliant with Industry Canada RSS 210 SUBPART INTENTIONAL RADIATORS FCC P...

Страница 12: ... For products sold in Europe the end of life procedures for battery operated electronic products must comply with the RoHS Directive 2002 95 EC the WEEE Directive 2002 96 EC and the Bat tery Directive 2006 66 EC These directives dictate how to dispose of the electronic and battery elements of the product after use For Protégé products sold in the UK only 3M has provided a col lection service This ...

Страница 13: ... not operate without power applied It ONLY detects gases while powered CAUTION Periodically test for correct operation of the system s alarm events by exposing the monitor to a targeted gas concentration above the high alarm set point CAUTION Verify the gas inlet ports are free of dirt and debris prior to use CAUTION Do not expose the monitor to severe mechanical or electrical shock Always conduct...

Страница 14: ...rosols Brake cleaner Bug repellents sprays Lubricants Lubricants Rust inhibitors Rust inhibitors Window glass cleaners Window glass cleaners Dish soaps Citrus based cleaners Alcohol based cleaners Hand sanitizers Anionic detergents WARNING REPLACE THE BATTERIES IMMEDIATELY WHEN THE DETECTOR EMITS A LOW BATTERY ALARM WARNING USE ONLY BATTERIES THAT ARE RECOMMENDED BY 3M WARNING DISPOSE OF USED LITH...

Страница 15: ...2 1 Button Definitions in Quick Reference for the operation button shown in the flow chart Figure 2 1 Power up Sequence Flow Chart WARNING THIS SECTION IS NOT A SUBSTITUTE FOR READING THE ENTIRE USER GUIDE ALL INDIVIDUALS WHO HAVE OR WILL HAVE RESPONSIBILITY OF USING OR SERVICING THE DEVICE MUST READ AND UNDERSTAND THE CONTENTS OF THE USER GUIDE PRIOR TO OPERATION FAILURE TO DO SO MAY RESULT IN SE...

Страница 16: ...perating The following flow chart provides a quick reference of the Protégé SG monitor s basic operation See Section 5 2 Operation Buttons for the detailed actions Refer to Table 2 1 Button Definitions in Quick Reference for the operation button shown in the flow chart below Figure 2 2 Basic Operating Flow Chart ...

Страница 17: ...tor s zero calibration See Section 5 2 Operation Buttons and Section 6 1 1 ZERO CALIBRATION for the detailed actions Refer to Table 2 1 Button Definitions in Quick Reference for the operation button shown in the flow chart below Figure 2 3 Zero Calibration Flow Chart WARNING MAKE SURE THE AREA IS FREE OF FLAMMABLE GASES DURING CALIBRATION FAILURE TO DO SO COULD RESULT IN SERIOUS INJURY OR DEATH ...

Страница 18: ...tor s span calibration See Section 5 2 Operation Buttons and Section 6 1 3 SPAN CALIBRATION for the detailed actions Refer to Table 2 1 Button Definitions in Quick Reference for the operation button shown in the flow chart below Figure 2 4 Span Calibration Flow Chart WARNING MAKE SURE THE AREA IS FREE OF FLAMMABLE GASES DURING CALIBRATION FAILURE TO DO SO COULD RESULT IN SERIOUS INJURY OR DEATH ...

Страница 19: ... SG monitor s bump test See Section 5 2 Operation Buttons and Section 6 1 4 BUMP TEST for the detailed actions Refer to Table 2 1 Button Definitions in Quick Reference for the operation button shown in the flow chart below Figure 2 5 Bump Test Flow Chart WARNING MAKE SURE THE AREA IS FREE OF FLAMMABLE GASES DURING BUMP TEST FAILURE TO DO SO COULD RESULT IN SERIOUS INJURY OR DEATH ...

Страница 20: ...3M PROTÉGÉ SG PORTABLE REUSABLE SINGLE GAS MONITOR QUICK REFERENCE 087 0065 Rev D February 2019 2 6 This page is intentionally left blank ...

Страница 21: ...connection allows quick parameter configuration monitor calibration and event log retrieval via the Portable Gas Detection PGD Application Software The list below shows all the available interchangeable smart sensor options for the Protégé SG mon itor CO Carbon Monoxide CO Carbon Monoxide H2 Resistant H2S Hydrogen Sulfide H2S Hydrogen Sulfide High Resolution O2 Oxygen O3 Ozone HCN Hydrogen Cyanide...

Страница 22: ...tor The monitor uses a special high viewing angle LCD designed to enhance the screen visibility see Figure 3 2 Monitor LCD During operation the LCD displays continuous data on gas concentra tions and alarm conditions Additionally you can monitor peak values of gas concentrations Table 3 1 Major Parts of the Monitor No Item No Item 1 Compliance LED Green 5 Buzzer 2 Side Alarm LEDs Red 6 IR Link 3 L...

Страница 23: ...tes Calibration Certificate Table 3 2 Monitor LCD Indicators No Description Mo Description 1 Battery Indicator Alarm Condition Icon 9 TWA Alarm Condition 2 Gas Type 10 STEL Alarm Condition 3 Heart Beat 11 LOW Alarm Condition 4 Gas Reading Or Real Time Clock 12 HIGH Alarm Condition 5 Gas Type 13 IRDA Is On data Transmission 6 Percentage By Lower Explosive Limit lel 14 Calibration Indication 7 Parts...

Страница 24: ...3M PROTÉGÉ SG PORTABLE REUSABLE SINGLE GAS MONITOR INTRODUCTION 087 0065 Rev D February 2019 3 4 This page is intentionally left blank ...

Страница 25: ...to open it and then double click the exe file to open it to the same folder 2 Double click the exe file to start the installation procedure 3 When prompted click Next to proceed with the installation see Figure 4 1 USB Driver Installa tion 1 of 4 through Figure 4 4 USB Driver Installation 4 of 4 Table 4 1 Installation Checklist Item Source Microsoft Net Framework Client V4 0 or above http msdn mic...

Страница 26: ...ABLE SINGLE GAS MONITOR CONFIGURATION AND SETUP 087 0065 Rev D February 2019 4 2 Figure 4 1 USB Driver Installation 1 of 4 4 Select I accept this agreement and then click Next to continue Figure 4 2 USB Driver Installation 2 of 4 ...

Страница 27: ...ure 4 3 USB Driver Installation 3 of 4 6 Once the driver has successfully installed click Finish to complete the installation see Figure 4 4 USB Driver Installation 4 of 4 Figure 4 4 USB Driver Installation 4 of 4 4 3 Connecting the Device to a PC NOTE This assumes that you have already installed the PGD application software and IR connect driver on your PC ...

Страница 28: ... 5 Connecting the IR Connect to a PC Figure 4 5 Connecting the IR Connect to a PC Figure 4 6 IRDA Indicator 3 Click on the PGD Application Software on the desktop The monitor is ready for a READ DEVICE command NOTE When face down ensure monitor is level or it will not be able to communicate with the IR programmer When face up check for the Infrared Data Transfer Icon to appear This confirms commun...

Страница 29: ...re Default Settings provide the default settings of the monitor NOTE The user can configure alarm setpoints and the firmware settings via the PGD Appli cation software Refer to the Portable Gas Detection PGD Application User Guide P N 087 0061 Table 4 2 Factory Default Alarm Setpoints Gas Range TWA STEL Low High Gas Type Value for CAL O2 0 30 N A N A 19 5 23 5 Zero air H2S Standard 0 500ppm 10ppm ...

Страница 30: ...O2 PH3 0 to 20 ppm 0 3 ppm 1 ppm 0 3 ppm 1 ppm 1 ppm PH3 SO2 0 to 150 ppm 2 ppm 5 ppm 2 ppm 5 ppm 10 ppm SO2 NH3 0 to 300 ppm 25 ppm 35 ppm 25 ppm 50 ppm 50 ppm NH3 Cl2 0 to 50 ppm 0 5 ppm 1 ppm 0 5 ppm 1 ppm 5 ppm Cl2 ClO2 0 to 5ppm 0 1 ppm 0 3 ppm 0 1 ppm 0 3 ppm 2 ppm Cl2 Table 4 3 Firmware Default Settings Options Range Factory Default Backlight Duration 2 255 seconds 30 seconds Calibration Pe...

Страница 31: ...and set the monitor parameter select the monitor that needs to be configured and dou ble click Or click Configure to display the configuration window 4 6 Sensor Information Perform the following steps to set the alarm setpoint 1 The monitor setting window displays see Figure 4 9 Sensor Information Window Figure 4 9 Sensor Information Window 2 Change the alarm set point 3 After completing all the c...

Страница 32: ...0065 Rev D February 2019 4 8 4 7 Monitor Configuration Click the Monitor Information tab to view and configure the monitor parameters see Figure 4 10 Monitor Information Window When completing the configurations click Set to save the settings Figure 4 10 Monitor Information Window ...

Страница 33: ...E DEVICE UNTIL IT HAS BEEN PROPERLY CALIBRATED FAILURE TO HEED THIS WARNING COULD RESULT IN DEATH OR INJURY WARNING IF MONITOR DOES NOT OPERATE IN THE MANNER DESCRIBED HERE DO NOT USE TAG IT OUT OF SERVICE FAILURE TO ENSURE IT IS PROPERLY OPERATING MAY RESULT IN SERIOUS INJURY OR DEATH WARNING YOU MUST FAMILIARIZE YOURSELF WITH THE ICONS IN BOTH THE NON ALARM AND ALARM STATES WARNING IF THE DISPLA...

Страница 34: ...Bump Test Indication 8 10 6 1 Ppm 16 Peak Gas Exposure NOTE The heart beat icon indicates that the monitor is in the gas reading mode and is working properly Table 5 2 Operation Buttons Function Button Action Description Turn On Left Press and hold the left but ton until the countdown is complete Activate the monitor During gas monitoring mode this action will start the bump test Clear Peak or TWA...

Страница 35: ...View Alarm Settings Right Press the right button when viewing peak or TWA value View high low STEL TWA and peak gas exposures Zero Calibration Press and hold right button in gas monitoring mode Initiate the calibration Cancel oper ation Press right button when operating Cancel the operation For example can cel the calibration or bump test Turn Off Both Press and hold both buttons until the countdo...

Страница 36: ...r s LCD using the Portable Gas Detection PGD Application Table 5 3 Self Test Sequence LCD Display Description System ON indicator Firmware Version Year Month and day Real Time The Alarm Settings screen s appear Note The number of Alarm Settings screens vary depending on the type of sensor installed in the device The SPAN Due Days screen appears The Bump Test Due Days screen appears The Calibration...

Страница 37: ...ther error codes appear during the operation see Section 6 3 Error Codes Table 5 4 Monitor Alerts and Alarms Descriptions LCD Description LCD Description Over Limit Exposure Alarm 3 fast LED flashes every 2 seconds 3 fast sounding beeps every 2 seconds 1 vibration every 2 sec onds ALARM and flash Negative Drift Alarm 1 LED flash every 2 sec onds 1 sounding beep every 2 seconds 1 vibration every 2 ...

Страница 38: ...y 2 sec onds STEL and gas reading flash Low Battery Alarm Battery icon flashes 1 beep 1 LED and 1 vibra tion every minute The device deactivates with the countdown depleted Bump Test and Calibration Due BUMP flashes CAL flashes Error Alarm EOX flashes 3 fast LED flashes every 2 seconds 3 fast sounding beeps every 2 seconds 1 vibration every 2 sec onds The device power off with the countdown Calibr...

Страница 39: ...e Gas Monitor OPERATION 087 0065 Rev D February 2019 5 7 NOTE When the Operation Buttons are held for three 3 seconds after the LCD goes blank the device attempts to power back on The timing parameter varies as set using the PGD Application ...

Страница 40: ...3M Protégé sg Portable reusable single Gas Monitor OPERATION 087 0065 Rev D February 2019 5 8 This page is intentionally left blank ...

Страница 41: ...entration of gas Sensors can lose sensitivity through normal degradation exposure to high gas concentra tions or sensor poisoning Accurate calibration can be achieved only if specific concentrations of the correct gases are used Bump Test Verifies the Span Calibration by subjecting the device to a known concentration of gas and verifies that the display reads PASS Table 6 1 Recommended Calibration...

Страница 42: ...on or wait for 60 seconds to go to span calibration Verify the device responds to target gas and activates 1 green LED flash every 5 seconds 1 sounding beep every 5 seconds and 1 vibration every 5 seconds Bump Test Daily When a sensor is exposed to a high concentra tion of gas When submerged in water When exposed to mechanical shock such as being dropped Refer to Section 6 1 4 BUMP TEST WARNING ON...

Страница 43: ... Only use the 3M approved gas to calibrate the monitor 1 Verify the concentration level of the target gas in the cylinder exceeds alarm settings of the mon itor refer to Section 4 4 Default Settings if needed 2 Attach the Regulator to the gas cylinder and verify cylinder pressure 3 Connect the tubing to both the 0 5LPM regulator on the gas cylinder and the calibration adapter Figure 6 2 SPAN Calib...

Страница 44: ...countdown see Figure 6 3 SPAN Calibration 6 Wait for the display to read SPAN PASS see Figure 6 3 SPAN Calibration 7 Verify the monitor responds to target gas green LED flashes once every 5 seconds beeps once every 5 seconds vibrates once every 5 seconds 8 Press the right button or wait 60 seconds to go to the Exit screen 9 Press the left button or wait for 60 seconds to return to normal reading s...

Страница 45: ... the left button to acknowledge the bump test and it will apply gas 5 Verify monitor responds to target gas and wait for the display to read PASS Verify the device does all of the following red LED flash once per second beeps once per second vibrates once per second 6 Press any button or wait 60 seconds to go to the normal reading screen 7 After an unsuccessful bump test the monitor LCD displays F...

Страница 46: ...and firmly pull the sensor cover apart Do not pull apart at an angle or by twisting 4 Pull the sensor straight out 5 Insert the replacement sensor using the Alignment System see Figure 6 5 Sensor Positions NOTE The torque of the screwdriver should be 3 to 4 inch pounds to loosen or tighten the screws NOTE Each sensor has a small PCB tab that extends beyond its diameter Ensure the tabs align with t...

Страница 47: ...65 Rev D February 2019 6 7 Figure 6 5 Sensor Positions 6 Reinstall the sensor cover with four 3 M2 5 X 12 screws and tighten the screws with the sequence shown in Figure 6 6 Screw Sequence Figure 6 6 Screw Sequence 1 2 Upper Alignment Tab Lower Alignment Tab 1st 3rd 2nd ...

Страница 48: ...S to remove the sensor before replacing battery 5 Replace with the new battery WARNING AFTER REPLACEMENT PERFORM A BUMP TEST AND CALIBRATION TO ENSURE PROPER OPERATION PRIOR TO PLACING BACK INTO SERVICE FAILURE TO FOLLOW THIS WARNING COULD LEAD TO SERIOUS INJURY OR DEATH WARNING DO NOT REPLACE BATTERY WHILE IN A HAZARDOUS LOCATION FAILURE TO FOLLOW THIS WARNING COULD LEAD TO SERIOUS INJURY OR DEAT...

Страница 49: ...BLE SINGLE GAS MONITOR MAINTENANCE 087 0065 Rev D February 2019 6 9 Figure 6 7 Lithium Battery Access Figure 6 8 Correct Polarity Orientation M2 5 x 10 Screws Sensor Cover Sensor Battery Battery Positive Terminal Negative Terminal ...

Страница 50: ...the screws with the sequence illustrated in Figure 6 6 Screw Sequence WARNING AFTER REPLACEMENT PERFORM A BUMP TEST AND CALIBRATION TO ENSURE PROPER OPERATION PRIOR TO PLACING BACK INTO SERVICE FAILURE TO FOLLOW THIS WARNING COULD LEAD TO SERIOUS INJURY OR DEATH WARNING DO NOT REPLACE FILTERS WHILE IN A HAZARDOUS LOCATION FAILURE TO FOLLOWING THIS WARNING COULD LEAD TO SERIOUS INJURY OR DEATH CAUT...

Страница 51: ...en delete the datalog through the PGD soft ware refer to PGD software user guide E05 Battery depleted Chargeable battery charge the battery un rechargeable battery replace the battery E06 Missing sensor wrong sen sor installation Reinstall the sensor or contact the service center E07 Sensor can not communi cate with Mipex sensor Re install sensor module and restart the monitor E08 Temperature sens...

Страница 52: ...3M PROTÉGÉ SG PORTABLE REUSABLE SINGLE GAS MONITOR MAINTENANCE 087 0065 Rev D February 2019 6 12 This page is intentionally left blank ...

Страница 53: ...Data Log 5 000 events Housing TPE Thermoplastic Elastomer Dimensions 4 2 H x 2 3 W x 0 9 D 106mmH x 58mmW x 23mmD LCD Size 0 7 W x 1 2 L 18mm x 30mm Weight 5 3oz 150g Operating Temperature Range 4 to 122 F 20 to 50 C Continuous long term operation Operating Humidity 0 to 95 RH non condensing Ingress Protection IP 65 67 Compliant NOTE These values are factory defaults and may be changed Table A 2 S...

Страница 54: ...C to 50 C Operating Humidity 15 to 90 RH non condensing HCN Hydrogen Cyanide Sensor Range 0 to 100 ppm Operating Temperature 4 F to 122 F 20 C to 50 C Operating Humidity 15 to 90 RH non condensing NO Nitric Oxide Sensor Range 0 to 350 ppm Operating Temperature 4 F to 122 F 20 C to 50 C Operating Humidity 15 to 90 RH non condensing NO2 Nitrogen Dioxide Sensor Range 0 to 100 ppm Operating Temperatur...

Страница 55: ...y 15 to 90 RH non condensing ClO2 Chlorine Dioxide Sensor Range 0 to 5 ppm Operating Temperature 4 F to 104 F 20 C to 50 C Operating Humidity 15 to 90 RH non condensing NH3 Ammonia Sensor Range 0 to 300 ppm Operating Temperature 4 F to 104 F 20 C to 50 C Operating Humidity 15 to 90 RH non condensing Note Due to a high cross sensitivity this sensor is unsuitable for use in areas where hydrogen sulf...

Страница 56: ...3M PROTÉGÉ SG PORTABLE REUSABLE SINGLE GAS MONITOR SPECIFICATIONS 087 0065 Rev D February 2019 A 4 This page is intentionally left blank ...

Страница 57: ...alibration Table B 1 Gas Interferences NOTE Table B 1 Gas Interferences does not show nor should it be implied that no addi tional interferences may occur These selectivity ratios are used as guides only They are not to be used as calibration factors The gas species actual cross sensitivities may vary from the values shown Interference Gas Sensor Types all values in ppm CO Single CO H 2 Resistant ...

Страница 58: ...opropyl Alcohol C3H7OH 0 002 0 Methanol CH3OH 0 Nitric Oxide NO 0 5 0 02 0 01 0 0 08 1 0 0 1 Nitorgen Dioxide NO2 0 25 0 01 0 2 4 0 3 1 0 6 2 Ozone O3 2 8 1 Phosphine PH3 1 Silane SiH4 0 9 Sulfur Dioxide SO2 0 25 0 001 0 1 0 0 3 0 0 0 2 1 Key Less than Approximate Grey highlights indicate calibration or surrogate calibration gases Interference Gas Sensor Types all values in ppm ...

Страница 59: ...Item Description Part Number Monitor Monitor RSG XLX0N X Accessories for Monitor Calibration Cap Adapter 8005276 Alligator Belt Clip 8005307 Battery ER14335 OmniCel 8005280 IR Programmer 093 0588 Internal Filter Assembly PSG 8005287 Manual Protégé SG Operation Mainte nance Manual 087 0065 Calibration Kit The spe cific Calibration Kit depends on the sensors in your device 4 gas Calibration Kit 0 5L...

Страница 60: ...ART NUMBERS Sensor EChem H2S 0 to 500 ppm 8005318 H2S 0 to 100 ppm High Resolution 8005320 CO Standard 0 to 1 000 ppm 8005321 CO 0 to 2000 ppm H2 Resistant 8005322 O2 0 to 30 8005324 SO2 0 to 150 ppm 8005325 PH3 0 to 20 ppm 8005326 O3 0 to 3 ppm 8005327 Cl2 0 to 50 ppm 8005328 ClO2 0 to 5 ppm 8005329 HCN 0 to 100 ppm 8005330 NO 0 to 350 ppm 8005331 NO2 0 to 100 ppm 8005332 NH3 0 to 300 ppm 8005333...

Страница 61: ...xed Gas Flame Detection 4055 Technology Forest Blvd The Woodlands TX 77381 United States Phone 1 713 559 9230 Email Detcon Service mmm com Europe Middle East Africa GMI Europe Gas Measurement Instruments Ltd Service Calibration Division 25 Cochran Close Crownhill Milton Keynes England MK8 0AJ Phone 44 0 1908 568 867 Fax 44 0 1908 261 056 Email gmiuk customersupport mmm com Asia Pacific 3M Gas Flam...

Страница 62: ...ÉGÉ SG PORTABLE REUSABLE SINGLE GAS MONITOR 087 0065 Rev D February 2019 D 2 Fixed Gas Flame Detection 4500 Technology Forrest Blvd The Woodlands TX 77381 Phone 713 559 9230 Web Site https gasdetection 3m com ...

Отзывы: