- 11 -
Cleaning the straight hand piece (straight nose cone) attachment (optional accessory)
Clean the straight hand piece preferably in an ultrasonic tank and rinse in distilled water.
Dry the straight hand piece.
Disinfect the straight hand piece with a cloth and mild disinfectant containing a neutral pH.
Seal the straight hand piece inside a sterilization pouch. If you are autoclaving more than 1
tip, place each straight hand piece in an individual bag. You may also use the tip holder that
comes with the ART-SP2 in place of the sterilization pouch.
Autoclave the straight hand piece .
Cleaning the Prophy Angle
(optional accessory)
The prophy angle should be cleaned to remove hair and debris approximately after every 10 dental
polishing.
1. Remove prophy angle from handpiece and rubber cup.
2. Remove head gears by turning knurled knob clockwise ( threads are reversed) turn to your right.
3. Clean with fine brush and lubricate head gears with mineral oil, Vaseline or any non-toxic
lubricant.
4. Place head gears back and tighten by turning Left or counterclockwise.
5. Normal polishing speeds are around 3,000 to 3,000 RPM, exceeding this may damage your
prophy angle.
Section V: Installation
5-A General Information
If the installation of your ART-SP2 system is performed by non-Bonart dealer personnel, check to
see that the requirements below are followed.
5-B Water Line Requirements
Incoming water supply line pressure to the Scaler must be 25 psi (172 kPa minimum) to 60
psi (414Kpa maximum). If your dental water supply line pressure is above 60 psi, install a
water regulator on the water supply line to your ART-SP2 unit.
After installing the water supply system, flush the water thoroughly prior to connecting to
the scaler.
Manually shut off the water valve of your dental water supply system prior to leaving the
office unoccupied and unattended.
5-C Electrical Requirements
Refer to Section X: Specifications