Bonart ART-SP2 Скачать руководство пользователя страница 1

 

 

ART-SP2 Scaler /Polisher Combo Unit 

Instruction Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bonart Co., Ltd.

 

 

 

Jorgen Kruuse A/S 

No.80, Wuxun St., Anle Dist.,   
Keelung City, 20446, Taiwan, R.O.C. 
Tel: 886-2-22983980   
Fax:886-2-22983981 
Email:[email protected] 

 

Havretoften 

4, 

DK-5550 

Langeskov, 

Denmark 

 

2460

 

COPYRIGHT©  2018 BONART CO., LTD.

 

 

 

Содержание ART-SP2

Страница 1: ...n Manual Bonart Co Ltd Jorgen Kruuse A S No 80 Wuxun St Anle Dist Keelung City 20446 Taiwan R O C Tel 886 2 22983980 Fax 886 2 22983981 Email generalinquiries bonartmed com Havretoften 4 DK 5550 Langeskov Denmark 2460 COPYRIGHT 2018 BONART CO LTD ...

Страница 2: ...vertent or deliberate publication BONART intends to enforce its rights to this manual under the copyright law Reproduction of the materials presented in this manual without the express written permission of BONART Co Ltd is prohibited The contents of this manual are subject to change without prior notice PROPERTY OF BONART CO LTD ALL RIGHTS RESERVED ART is a registered trademark of BONART Co Ltd i...

Страница 3: ...perly trained personnel Only qualified personnel shall perform any adjustments maintenance and repair of the equipment NOTE CAUTION AND WARNING STATEMENTS NOTE Provide tips and advice CAUTION Provide correct operating or maintenance procedures WARNING Alert user of danger of injury or damage SYMBOLS Complies with MDD 93 42 EEC BF equipment Grounding terminal Attention Please read instructions ...

Страница 4: ...kers Patients with pacemakers should avoid treatment with the ART SP2 Studies have shown that electronic appliances including razors hair dryers microwave ovens TV receptors and other medical equipment such as the ART SP2 may interfere with the performance of pacemakers For more information on this subject please refer to the following articles Advances in Cardiac Pacemaker The New York Academy of...

Страница 5: ... or unstable surface may degrade performance or damage the device Do not modify the device Modifications to the device will violate safety codes and invalidate the warranty while endangering the patient and operator Do not place heavy objects on top of the power cord and keep the power cord away from intense heat Unplug the device and call Bonart for service and instructions if you observe any abn...

Страница 6: ...noise and heat and efficiently reduces the amount of cooling water required during treatment This electric mechanical feature provides greater efficiency than traditional magnetostrictive systems Features Scaler and polisher in one system Automatic phase lock and power control Automatic feedback control Micro motor speed control of 2 000 to 30 000 RPMS Supplies Replacement Part To order supplies o...

Страница 7: ...on II Contraindications and Warnings 7 Section III Precautions 7 Section IV Infection Control 9 Section V Installation 11 Section VI General Description and Information of Parts 16 Section VII Techniques 18 Section VIII System Maintenance 19 Section IX Troubleshooting 20 Section X Specifications 21 Section XI Disposal 21 ...

Страница 8: ... Do not immerse the ART SP2 in water or liquid If the ART SP2 has water damage return the machine to Bonart for servicing Do not modify the ART SP2 Modifications will invalidate the warranty on the machine violate safety codes and endanger the patient and operator Section Ⅲ Precautions 3 A Precautions for the ART SP2 and all Ultrasonic Scaler Systems Ensure that an adequate amount of water flows t...

Страница 9: ...d tips BONART piezo tips are made of stainless steel Avoid direct contact with teeth to prevent damage to the enamel We strongly advise practicing on models and aluminum plates prior to operating on a live subject with the ART SP2 Avoid direct contact of piezo tips with the subject s lips cheeks and tongue Wipe accretions from the tooth using the long axis of the scaler tip DO NOT gouge the tooth ...

Страница 10: ...and piece 2 Tips 3 Metal Wrench 4 Micro motor 5 Straight hand piece Straight nose cone optional accessory 6 Prophy angle unless disposable optional accessory The hand piece tips wrench straight handpiece and prophy angle can be autoclaved and can withstand a maximum temperature of 135 or 275 for a maximum of 3 minutes After completing the sterilization cycle air dry for about 20 30 minutes before ...

Страница 11: ...ining a neutral pH Seal the tip or wrench inside a sterilization pouch If you are autoclaving more than 1 tip place each tip or wrench in an individual bag You may also use the tip holder that comes with the ART SP2 in place of the sterilization pouch Autoclave the tip or wrench WARNING Before starting the sterilization cycle make sure that the inside of the tip is completely dry by blowing air th...

Страница 12: ... are reversed turn to your right 3 Clean with fine brush and lubricate head gears with mineral oil Vaseline or any non toxic lubricant 4 Place head gears back and tighten by turning Left or counterclockwise 5 Normal polishing speeds are around 3 000 to 3 000 RPM exceeding this may damage your prophy angle Section V Installation 5 A General Information If the installation of your ART SP2 system is ...

Страница 13: ...d Wrench 1 8 ART BS1 Universal Pointed Tip ART BS2 Chisel Head Flat Tip ART BS3 Round Flat Tip 1 1 1 9 Micro motor 1 10 Micro motor seat Optional 11 Straight Hand Piece Optional NOTE Other piezo tips are available and can be purchased separately through dealers Included items are subject to change depending on location NOTE While unpacking check your ART SP2 unit for any damages If any damages are...

Страница 14: ...ne Connection Push the blue water tube hose over the stainless steel receptacle on the back of the unit Then tighten the screw over it Connect the quick connector to the water supply line The ART SP2 standard package comes with a male quick connector If a female quick connector is necessary please contact a Bonart authorized dealer to purchase it Inspect all connections for leaks To remove the wat...

Страница 15: ...ro motor by following the steps below 1 Connect the straight hand piece and the micro motor by pushing the straight hand piece over the rod of the micro motor It should snap into place Figure 3 2 Attach the prophy angle to the straight hand piece by following the steps below A Unlock the prophy angle from the straight hand piece by turning the locking ring clockwise This should open the lock Note ...

Страница 16: ...y angle and put paste in the cup to begin polishing 2 Press the Micro motor mode button to switch from scaling to polishing mode 3 Adjust the Oscillating dial to a low speed 2 000 rpm 4 Press on the foot switch to engage the polishing hand piece and begin polishing 5 Add paste as needed Assembling the Scaler Unit 1 Ensure that the power switch is set to off If plugged in the power indicator light ...

Страница 17: ... are driven by the electrical signal from the Main unit The ART SP2 unit is compatible with Bonart BS type piezo tips and other brands of piezo tips with a comparable thread Do not use Bonart BS type piezo tips with the ART SP2 3 Power button ON OFF The power button is used to turn the ART SP2 main unit on and off When the power is on the LED indicator should be lit When turned off the LED power i...

Страница 18: ...trol the power intensity When turned clockwise the power intensity increases When turned counter clockwise the power intensity decreases 7 Micro Motor Socket The micro motor socket is where you connect the micro motor to the main unit It is located on the back of the main unit at the bottom left 8 Footswitch Socket The foot switch socket is where you connect the foot switch to the main unit It is ...

Страница 19: ...e over a sink or drain with a tip installed Verify that water is reaching the tip 6 Check your Bonart piezo tip for wear and replace as needed 7 Use the foot control to regulate water flow to the hand piece Hold down the foot control to allow water to flow through Release the foot control to stop 8 Adjust the water flow to the desired level using the water adjustment knob Increase the water flow b...

Страница 20: ...hould be rinsed with clean water for about 20 to 30 seconds 2 Inspect the hand piece cable daily to ensure it is in good condition 3 For sterilization of the hand piece please refer to Section IV 8 B TIP maintenance Worn out tips can adversely affect performance resulting in insufficient power and vibration Check tips regularly for wear and tear and replace as necessary 8 C Main Unit Maintenance T...

Страница 21: ...ter supply is not connected to the unit properly Check your water supply for sufficient flow and ensure that the water tube is properly secured to the back of the unit The filter may be clogged Clean or replace the filter The foot switch may be broken defective NOTE only applies if there is no vibration as well Reposition the foot switch and cable to see if there might be a short in the wire Repla...

Страница 22: ...l Requirement for Safety EN60601 1 2 Medical Electrical Equipment Electromagnetic Compatibility Requirements and Tests Specification Power supply 115V 5 50 60Hz 160VA 230V 5 50 60Hz 160VA Working frequency 26KHz 32KHz Polisher speed 2000 30000 R P M Water supply 25 60 PSIG 172 414KPa Dimension 24 5cm L x 22 5cm W x 8 5cm H Weight 3 5 Kg including the handpiece Hand piece Cable 250 cm Footswitch Ca...

Отзывы: