AKS17 ins
tallation and oper
ation ins
truction manual (2021/06/16)
6/14
System requirements
The programming device can be connected to a Windows PC with a USB cable and operated with the
dedicated software. The software requires Windows Vista or later (32/64 Bit).
Before the programming device (Programmer AKS17) can be connected the specific driver must be installed.
Copy the software to a local directory on your Windows PC. To install the driver execute the specific program.
USB_Adapter_00052040_Driver.exe Connecting the device (for the calibration setup)
Mount the sensing head correctly per mechanical specifications. It is required to place the sensing head
within the allowed mechanical tolerances (maximum displacement), see technical data sheet.
Both status LEDs (green and red) on the AKS17 sensing head will light up now if all devices are mounted
correctly. The pictures below show the final stage of all connected cables and the programming device.
Programming software
Copy the programming software (AKS16 and AKS17 Calibration GUI) to a local directory or use the “Setup.
msi” to install the software and the driver. By default the installation directory can be found at “Program
Files”.