BPAP System User Manual
5 / 57
IMPORTANT TIPS!
• An irregular sleep schedule, alcohol consumption, obesity, sleeping pills, or sedatives may
aggravate your symptoms.
• Please use a mask which meets ISO 17510: 2015.
5. Specifications
Device Size
Dimensions (L x W x H): 265 mm × 145 mm × 114 mm
Weight: 1.7 kg
Water capacity:
To maximum fill line 360 mL
Product Use, Transport and Storage
Operation
Transport and Storage
Temperature: 5°C to 35°C (41˚F to 95˚F) -25°C to 70°C (-13˚F to 158˚F)
Humidity: 15% to 93% Non-condensing
15% to 93% Non-condensing
Atmospheric Pressure: 760
~
1060 hPa
760
~
1060 hPa
Heated Humidifier
Humidifier Settings: Off, Auto, 1 to 5 (95˚F to 154.4˚F / 35°C to 68°C)
Humidifier Output: No less than 15 mg H
2
O/L
Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity
Maximum Operating Pressure: 40 hPa
Pressure Drop with Humidifier: < 0.4 hPa at 60 LPM flow
Maximum Delivered Gas Temperature: ≤ 43°C
Cellular Module
Receiver Frequency Band: 850/900/1800/1900 MHz
FCCID: XMR201202M35
Max RF power output: 33.0 dBm
WiFi Kit
FCCID: 2ACSVHF-LPT270
Mode of Operation
Continuous
Work Mode
CPAP, AutoCPAP, S, AutoS, S/T, T
SD Card
The SD card can record patient data and fault information
AC Power Consumption
100 - 240 V
~
, 50 / 60 Hz, 2.5 A Max
100 - 240 V
~
, 50 / 60 Hz, 2 A Max
Содержание G3 B20A