BPAP System User Manual
56 / 57
Recommended separation distances between RF wireless communications
equipment
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between RF wireless
communications equipment and the device as recommended below, according to the maximum
output power of the communications equipment.
Frequency
MHz
Maximum
Power
W
Distance IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
385
1.8
0.3
27
27
RF wireless communications
equipment should be used
no closer to any part of the
device, including cables,
than the recommended
separation
distance
calculated from the equation
applicable to the frequency
of the transmitter.
Recommended
separation distance
Where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF
transmitter,
as
determined
by
an
electromagnetic
site
survey, should be less than
the compliance level in
each frequency range.
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
450
2
0.3
28
28
710
0.2
0.3
9
9
745
780
810
2
0.3
28
28
870
930
1720
2
0.3
28
28
1845
1970
2450
2
0.3
28
28
5240
0.2
0.3
9
9
5500
5785
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected