BPAP System User Manual
8 / 57
Wavelengths
Red: 663 nanometers
Infrared: 890 nanometers
Maximal Optical Output Power
Less than 1.5 mw maximum average.
Air Tubing
Air tubing
Length
Inner diameter
Tubing
6 ft. (1.83 m)
19 mm
Heated Tubing
6 ft. (1.83 m)
19 mm
The Form and the Dimensions of the Patient Connection Port
The 22 mm conical air outlet complies with ISO 5356-1.
PM2.5 Filter
Efficiency: > 90% for 2.5 micron dust
6. Available Therapies
The device delivers the following therapies:
CPAP
–
Delivers Continuous Positive Airway Pressure; CPAP maintains a constant level of
pressure throughout the breathing cycle. If your health care professional has prescribed
ramp for you, you can turn
the Knob
to reduce the pressure and then gradually
increase the pressure to the therapeutic pressure setting so that you can fall asleep more
comfortably.
AutoCPAP
–
Delivers CPAP therapy and provides an air pressure no less than the prescribed
one based on the patient’s needs.
S
–
A bi-level mode which responds to both your inhalation and exhalation by increasing
pressure when you start to inhale and decreasing pressure when you start to exhale. There is
no automatic delivery of breathing gas if you do not inhale. IPAP (Inspiratory Positive Airway
Pressure) and EPAP (Expiratory Positive Airway Pressure) are preset by a home care provider.
AutoS
–
A bi-level mode which responds to both your inhalation and exhalation. The
differential pressure of IPAP and EPAP are presetted by a home care provider. While working
in auto mode, the device will automatically adjust the IPAP and EPAP if it detects a sleep
apnea.
T
–
A bi-level mode in which the device automatically starts inhalation and exhalation, and
automatically controls the time of inhalation and that of exhalation according to the preset
parameter.
S/T
–
A bi-level mode which responds to both your inhalation and exhalation by increasing
pressure when you start to inhale and decreasing pressure when you start to exhale. If you
do not start inhaling within a set time, the device will automatically start the process of
inhalation. When the device starts the process of inhalation, it controls the time of inhalation
and automatically decreases the pressure for exhalation within a set time.