BPAP System User Manual
40 / 57
High RR!!
(only applies to
G3 B25VT, G3
B30VT, G3 B30SV
and G3 LAB)
Middle
Priority
Alarm
When the airflow is on, an audible alarm
will sound if the respiratory rate exceeds
the alarm limit; the screen will display
―
High RR!!
‖.
Setting range:
Off, the setting value of
Low RR
~
80 BPM, in 1 BPM increments,
the default setting is ―
40 BPM
‖.
Note:
This function is avaliable under the
work mode of S/T or T.
Humidifier
Failure!!
Middle
Priority
Alarm
When humidifier is applied, an audible
alarm will sound when the humidifier fails
to work in 10 minutes; the screen will
display ―
Humidifier Failure!!
‖.
Please change
filter!
Low
Priority
Alarm
When the Filter Alarm feature is enabled,
an audible alarm will sound if the preset
replacement time is reached but the air
filter is not replaced; the screen will
display ―
Please change filter!
‖. The
default setting is ―
Off
‖.
Please replace
tubing!
Low
Priority
Alarm
When the tubing Alarm feature is
enabled, an audible alarm will sound if
the preset replacement time is reached
but the tubing is not replaced; the screen
will display ―
Please replace tubing!
‖.
Please replace
mask!
Low
Priority
Alarm
When the Mask Alarm feature is enabled,
an audible alarm will sound if the preset
replacement time is reached but the mask
is not replaced; the screen will display
―
Please replace mask!
‖.
SD Card Full!
Low
Priority
Alarm
The screen will display ―
SD Card Full!
‖ if
the SD card has reached its maximum
capacity.
Reinsert SD card!
Low
Priority
Alarm
The screen will display ―
Reinsert SD
card!
‖ if the SD card fails to work.
SpO
2
faults!
Low
Priority
Technology
Alarm
Disconnect the PULSE OXIMETER PROBE
from the PULSE OXIMETER EQUIPMENT
and place it in series with a circuit with
which each PULSE OXIMETER PROBE
wire can be opened or shorted to any
other PULSE OXIMETER PROBE wire. The
screen will display
“SpO
2
faults!”
.
Note: the delay time of alarming system of the device is no more than 1 s.