BPAP System User Manual
46 / 57
18.8 Disinfection
Generally speaking, if you have strictly followed the above cleaning instructions, you do not
have to disinfect the device and / or the water chamber. If the device is contaminated or
used in clinical trials, you can purchase disinfectants from a home medical equipment
company to disinfect the device.
Disinfection of the Water Chamber:
Prior to disinfection, clean the water chamber in accordance with Section 18.3 ―Cleaning the
Water Chamber‖. The disinfection methods is as follows:
(1) Heat disinfection: Soak the water chamber in tap water at 75
°C
± 2
°C
for 30 minutes for
disinfection.
(2) Use mild disinfectants.
Disinfection of the SpO
2
Probe:
For details, refer to the disinfection instructions in the user manual for the SpO
2
Kit.
WARNINGS!
• After disinfection, rinse any disinfected component in clean water thoroughly, especially
components in close contact with the patient such as the mask, headgear, and tubing, so as
to prevent disinfectant residuals from damaging the skin or respiratory tract or causing
allergies.
• Sterilization of the device and its components other than that is recommended is not
permitted.
• In order to prevent cross-infection of patients or contamination of equipment, BSF
(Breathing System Filter) that meets ISO 23328-1:2003 and ISO 23328-2:2002 standards
and has medical device registration certificates can be used.
(1) Different patients need to replace a new BSF before using this equipment.
(2) When using the BSF, please install and operate it according to the instructions of the BSF,
and pay attention to adjust the output pressure setting of the device according to resistance
of the BSF to ensure delivering normal treatment pressure.
(3) Atomization or humidification will increase the resistance of the BSF. The operator must
often monitor the resistance increase and blockage of the BSF to ensure delivering normal
treatment pressure.
• If you use ozone or other cleaning and disinfection methods that are not recommended by
BMC, BMC will not be able to verify the safety or performance of the equipment.
CAUTIONS!
• Disinfectants tend to damage the materials and reduce the life of components. Try to select
the appropriate disinfectant, and follow the disinfectant manufacturer's instructions and
recommendations.
• After disinfection, check the disinfected component for any signs of damage. Replace any
damaged component immediately.