BPAP System User Manual
43 / 57
(176°F). High temperatures could reduce product life.
• Do not immerse the device in any fluids.
18.1 Cleaning the Mask and Headgear
For details, refer to the cleaning instructions in the user manual for the mask.
18.2 Cleaning the
SpO
2
Kit
For details, refer to the cleaning instructions in the user manual for the SpO
2
Kit.
18.3 Cleaning the Water Chamber
(1)
Opening the Water Chamber:
Open the cap of the water chamber, as shown in Fig.
18-1.
Fig. 18-1
(2)
Cleaning the Water Chamber:
You may also clean the water chamber with a soft cloth
which does not scratch the water chamber (dip the soft cloth in liquid
soap if necessary),
rinse it thoroughly, and then wipe it dry with a soft cloth.
(3)
Putting the Water Chamber back
according to instructions in 13.1.3.
WARNINGS!
• Emptying and cleaning the water chamber daily will help prevent mold and bacteria
growth.
• Allow the water in the chamber to cool down to room temperature before removing it from
the device.
CAUTIONS!
• Clean the water chamber only after the water in it cools. Make sure that no water enters
the device.
• After cleaning, rinse the water chamber throughly in clean water to make sure that no soap
residue is left; then wipe it dry with a lint-free cloth, so as to prevent calcareous
accumulations.
• Check the water chamber for any leak or damage. Replace the water chamber if there is
any damage.
• It is recommended to do daily cleaning of the water chamber.
18.4 Cleaning the Transfer Box
(1)
Removing the Transfer Box:
First remove the water chamber from the device, then
remove the transfer box, as shown in Fig. 18-2.