Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
user of the device should make sure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electrostatic discharge
(ESD) IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Electrical fast transient /
burst IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input / output lines
±2 kV for power supply lines
±1 kV for input / output lines
Surge IEC 61000-4-5
±1 kV line (s) to line (s)
±2 kV common mode
±1 kV line (s) to line (s)
±2 kV common mode
Voltage
dips,
short
interruptions
and
voltage variations on
power supply input lines
IEC 61000-4-11
0%
U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0%
U
T
; 1 cycle
70%
U
T
; 25 / 30 cycle
At 0°
0%
U
T
; 250 / 300 cycle
0%
U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0%
U
T
; 1 cycle
70%
U
T
; 25 / 30 cycle
At 0°
0%
U
T
; 250 / 300 cycle
Power frequency (50 /
60 Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Note:
U
T
is the AC mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
user of the device should make sure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Conducted RF
IEC 61000-4-6
3 V 0.15 MHz
~
80 MHz
6 V in ISM and amateur radio
bands between 0.15 MHz and
80 MHz
3 V 0.15 MHz
~
80 MHz
6 V in ISM and amateur radio
bands between 0.15 MHz and 80
MHz
Radiated RF
IEC 61000-4-3
10 V/m 80 MHz to 2.7 GHz
10 V/m 80 MHz to 2.7 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applied.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular /
cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
device is used exceeds the applicable RF compliance level above, the device should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 10
V/m.