72
2.
Press
button two times until RS-485 is blinking and press
to configure the settings for RS-485 remote
communication. The following display will be shown:
3.
Use
keys to select between each settings and press
to configure the corresponding setting.
4.
To change the parameter settings, use the rotary dial.
5.
Press
to save each setting and the display will return to the
RS485 menu.
6.
Below lists the options that can be changed for each setting:
Baud rate: 9600, 19200, 38400, 57600
Parity/Data bits: None/8 bits, Even/8 bits, Odd/8 bits, Even/7
bits, Odd/7 bits
Stop bit: 1, 2
Address: 1 to 50
Chain: On, Off
Note: The default is 9600, None/8 bits, 1, Address = 1, Off.
7.
For each power supply that you want to control, provide a
different address. For example, if you have three power
supplies to control, set the first supply to address 1, second
supply to address 2, and third supply to address 3. If two or
more supplies have the same address, the RS485 protocols will
create communication errors.