
103
Figure 5.7 – CC Calibration of External Current Diagram
2.
Set the DMM to a DC current measurement. Select the EXT_I_CC
parameter in the CAL menu and press
. The following will
be displayed:
3.
Input the current reading from the DMM into the METER
parameter by using the numeric keypad and press
.