97
4.
Repeat this procedure by entering the meter readings for points
2 – 4.
5.3
OVP Calibration
The overvoltage protection (OVP) calibration can be executed right after
the voltage calibration or the user can select OVP in the CAL menu.
Follow the steps below to perform the overvoltage protection
calibration:
1.
Disconnect any load from the output terminal of the power
supply.
2.
Select the OVP parameter in the CAL menu and press
. The
following will be displayed:
3.
Press
to run the OVP calibration or press
to abort. The
OVP calibration is an automatic process that will take about 10
to 20 seconds.
4.
After the calibration is finished, the following message will be
displayed: