background image

permitted. The HOLDOFF control allows this period to be
extended by a variable amount if desired.

This control is usually set to the MIN position (fully

counterclockwise) because no additional holdoff period is
necessary. The HOLDOFF control is useful when a com-
plex series of pulses appear periodically such as in Fig. 4B.
Improper sync may produce a double image as in Fig. 4A.
Such a display could be synchronized with the VAR
SWEEP 
control, but this is impractical because time meas-
urements are then uncalibrated. An alternate method of
synchronizing the display is with the HOLDOFF control.
The sweep speed remains the same, but the triggering of the
next sweep is “held off” for the duration selected by the
HOLDOFF control. Turn the HOLDOFF control clock-
wise from the MIN position until the sweep starts at the
same point of the waveform each time.

MAGNIFIED SWEEP OPERATION

Since merely shortening the sweep time to magnify a

portion of an observed waveform can result in the desired
portion disappearing off the screen, magnified display
should be performed using magnified sweep.

Using the

POSition control, move the desired portion

of waveform to the center of the CRT. Pull out the PULL X10
knob to magnify the display ten times. For this type of display
the sweep time is the Main Time Base TIME/DIV control
setting divided by 10. Rotation of the

POSition control can

then be used to select the desired portion of the waveforms.

X

Y OPERATION

X

operation permits the oscilloscope to perform many

measurements not possible with conventional sweep opera-
tion. The CRT display becomes an electronic graph of two
instantaneous voltages. The display may be a direct com-
parison of the two voltages such as stereoscope display of
stereo signal outputs. However, the X

mode can be used

to graph almost any dynamic characteristic if a transducer is
used to change the characteristic (frequency, temperature,
velocity, etc.) into a voltage. One common application is fre-
quency response measurements, where the Y axis corresponds to
signal amplitude and the X axis corresponds to frequency.

1.

Set the SWEEP MODE switch to the X

position.

Set the Trigger Source an d VERTical MODE
switches to X

Y.

2.

In this mode, channel 1 becomes the X axis input and

channel 2 becomes the Y axis input. The X and Y
positions are now adjusted using the

POSition and

the channel 2

POSition controls, respectively.

3.

Adjust the amount of vertical (Y axis) deflection with

the CH 2 VOLTS/DIV and VARIABLE controls.

4.

Adjust the amount of horizontal (X axis) deflection

with the CH 1 VOLTS/DIV and VARIABLE con-
trols.

VIDEO SIGNAL OBSERVATION

Setting the COUPLING switch to the TV-H or TV-V

position permits selection of horizontal or vertical sync
pulses for sweep triggering when viewing composite video
waveforms.

When the TV-H mode is selected, horizontal sync pulses

are selected as triggers to permit viewing of horizontal lines
of video. A sweep time of about 10

µ

s/div is appropriate for

displaying lines of video. The VAR SWEEP control can be
set to display the exact number of waveforms desired.

When the TV-V mode is selected, vertical sync pulses are

selected as triggers to permit viewing of vertical fields and
frames of video. A sweep time of 2 ms/div is appropriate for
viewing fields of video and 5 ms/div for complete frames
(two interlaced fields) of video.

At most points of measurement, a composite video signal

is of the (

) polarity, that is, the sync pulses are negative and

the video is positive. In this case, use (

) SLOPE. If the

waveform is taken at a circuit point where the video wave-
form is inverted, the sync pulses are positive and the video
is negative. In this case, use (+) SLOPE.

DELAYED SWEEP OPERATION

(Refer to Fig. 4)

Delayed sweep operation is achieved by use of both the

main sweep and the delayed sweep and allows any portion
of a waveform to be magnified for observation. Unlike X10
magnification, delayed sweep allows selectable steps of
magnification.

1.

Set the Sweep Mode switch to the MAIN position and

adjust the oscilloscope for a normal display.

2.

Set the Sweep Mode switch to the MIX position. The

display will show the main sweep on the left portion
(representing the MAIN Time Base control setting)
and the delayed sweep on the right portion (repre-
senting the DELAY Time Base control setting). The
MAIN Time Base portion of the trace usually will be
brighter than the delayed time base portion. Fig. 4
shows a typical display for the MIX display mode.

OPERATING INSTRUCTIONS

15

Содержание 2190B

Страница 1: ...100 MHz DUAL TRACE OSCILLOSCOPE BK PRECISION INSTRUCTION MANUAL MODEL 2190B BK PRECISION ...

Страница 2: ...et 4 Use the time proven one hand in the pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return path 5 When using a probe touch only the insulated portion Never touch the exposed tip portion 6 When testing ac powered equipment remember that ac line voltage is usually present on some power input c...

Страница 3: ...l for further precautionary information This symbol appears in the manual where the corresponding information is given Instruction Manual for Model 2190B 100 MHz Dual Trace Oscilloscope 2000 B K Precision Corp 1031 Segovia Circle Placentia CA 92870 BK PRECISION ...

Страница 4: ...ing Procedure 12 Single Trace Display 12 Dual Trace Display 12 Triggering 13 Magnified Sweep Operation 15 OPERATING INSTRUCTIONS Continued X Y Operation 15 Video Signal Observation 15 Delayed Sweep Operation 15 Applications Guidebook 16 MAINTENANCE 17 Fuse Replacement 17 Line Voltage Selection 17 Periodic Adjustments 17 Calibration Check 18 Instrument Repair Service 18 APPENDIX 19 Important Consid...

Страница 5: ... with X10 MAG assures high frequencies and short duration pulses are displayed with high resolution VERTICAL FEATURES High Sensitivity 5 mV div sensitivity for full bandwidth High sen sitivity 1 mV div and 2 mV div using PULL X5 gain control Calibrated Voltage Measurements Accurate voltage measurements 3 on 10 cali brated ranges from 5 mV div to 5 V div Vertical gain fully adjustable between calib...

Страница 6: ...its stable observation of complex pulse trains OTHER FEATURES Bandwidth Limiter Selectable full bandwidth of 100 mHz or limited band width of 20 MHz to filter high frequency noise from low frequency waveforms X Y Operation Channel 1 can be applied as horizontal deflection X axis while channel 2 provides vertical deflection Y axis Built in Probe Adjust Square Wave A 2 V p p 1 kHz square wave genera...

Страница 7: ...proximately 500 kHz Polarity Reversal CH 2 invert Maximum Input Voltage 400 V dc ac peak HORIZONTAL AMPLIFIER Input through channel 1 input X Y mode CH 1 X axis CH 2 Y axis Sensitivity Same as vertical channel 2 Input Impedance Same as vertical channel 2 Frequency Response DC to 2 MHz 3 dB X Y Phase Difference 3 or less at 100 kHz Maximum Input Voltage Same as vertical channel 1 SWEEP SYSTEM Opera...

Страница 8: ...nominal CH 2 Y Output Output Voltage 50 mV div nominal into 50 ohm load Output Impedance Approximately 50 ohms Frequency Response 20 Hz to 100 MHz 3 dB Intensity Modulation Input Signal TTL level intensity increasing with more positive levels decreased intensity with more negative levels Input Impedance Approximately 50 kΩ Usable Frequency Range DC to 5 MHz Maximum Input Voltage 30 V dc ac peak Po...

Страница 9: ...ical calibration VERTICAL CONTROLS 9 VERTical MODE Switch Selects vertical display mode Four position lever switch with the following positions CH1 Displays the channel 1 signal by itself CH2 X Y CH2 displays the channel 2 signal by itself X Y used in conjunction with the X Y control and Trigger SOURCE switch to enable X Y display mode DUAL Displays the channel 1 and channel 2 signals simul taneou...

Страница 10: ...hen it is pushed in the two sweeps are alternately displayed one after the other normally used at higher sweep speeds 15 CH2 POSition PULL INVert Control POSition Rotation adjusts vertical position of channel 2 trace In X Y operation rotation adjusts vertical position of X Y display PULL INVert When pushed in the polarity of the channel 2 signal is normal When pulled out the polarity of the channe...

Страница 11: ...e delayed sweep X Y Used with the VERTical MODE switch and Trig ger SOURCE switch to select X Y operating mode The channel 1 input becomes the X axis and the channel 2 input becomes the Y axis Trigger source and coupling are disabled in this mode TRIGGERING CONTROLS 26 HOLDOFF PULL NORM TRIG Control HOLDOFF Rotation adjusts holdoff time trigger inhibit period beyond sweep duration When control is ...

Страница 12: ...ts a positive going slope and the out position selectsanegative going slopeas triggeringpoint for main sweep 30 EXTernal TRIGger Jack External trigger input for single and dual trace operation REAR PANEL CONTROLS not shown 31 Fuse Holder Line Voltage Selector Contains fuse and selects line voltage 32 Power Cord Receptacle 33 CH 2 Y SIGNAL OUTPUT Jack Output terminal where sample of channel 2 signa...

Страница 13: ...veform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never allow a small spot of high brilliance to remain stationary on the screen for more than a few seconds The screen may become permanently burned A spot will occur when the scope is set up for X Y operation and no signal is applied Either reduce the intensity so the spot i...

Страница 14: ...in Time Base control to 1 mS div Sweep Mode switch to MAIN 2 Press the red POWER pushbutton 3 A trace should appear on the CRT Adjust the trace brightness with the INTENSITY control and the trace sharpness with the FOCUS control NOTE When viewing audio and low frequency waveforms below 20 MHz a better wave form may be obtained by engaging the 20 MHz 100 MHz pushbutton This limits the bandwidth to ...

Страница 15: ...tain a stable jitter free dis play in single trace or dual trace operation The proper settings depend upon the type of waveforms being observed and the type of measurement desired An explanation of the various controls which affect synchronization is given to help you select the proper setting over a wide range of conditions AUTO or NORM Triggering 1 In the AUTO triggering mode PULL NORM TRIG swit...

Страница 16: ... direction counterclockwise the triggering threshold shifts to a more negative value When the control is centered the threshold level is set at the approximate average of the signal used as the triggering source Proper adjustment of this control usually synchro nizes the display The TRIG LEVEL control adjusts the start of the sweep to almost any desired point on a waveform On sine wave signals the...

Страница 17: ...he Trigger Source and VERTical MODE switches to X Y 2 In this mode channel 1 becomes the X axis input and channel 2 becomes the Y axis input The X and Y positions are now adjusted using the POSition and the channel 2 POSition controls respectively 3 Adjust the amount of vertical Y axis deflection with the CH 2 VOLTS DIV and VARIABLE controls 4 Adjust the amount of horizontal X axis deflection with...

Страница 18: ...ase control Because of this the oscilloscope automat ically prevents electrically the DELAY Time Base from being set to a slower sweep speed than the MAIN Time Base For example if the MAIN TimeBase is set to 0 1 ms div the slowest possibleDELAY Time Base sweep speed is also 0 1 ms div even if the control is set slower APPLICATIONS GUIDEBOOK B K Precision offers a Guidebook to Oscilloscopes which d...

Страница 19: ...le Pry fuseholder away from receptacle 3 When reinstalling fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To select the desired line voltage simply insert the fuse and fuse holder so that the appropriate voltage is pointed to by the arrow Be sure to use the proper value fuse see label on rear panel PERIOD...

Страница 20: ...d only as a general check of calibration accuracy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is required for calibration adjustments The CAL signal should not be used as a time base standard INSTRUMENT REPAIR SERVICE Because of the specialized skills and test equipment re quired for instrument repair...

Страница 21: ...p speed and using X10 magnification For the Model 2190B this sweep rate is 2 ns div A rise time of less than about five divisions at this sweep speed should be calculated Calculated Measurements For observed rise times of less than 11 ns the pulse rise time should be caluclated to eliminate the error introduced by the cascaded oscilloscope rise time Calculate pulse rise time as follows Tpulse Limi...

Страница 22: ...amages so the above limitation or exclusion may not apply to you This warranty gives you specific rights and you may also have other rights which vary from state to state SERVICE INFORMATION Warranty Service Please return the product in the original packaging with dated proof of purchase to the address below Clearly state in writing the performance problem and return any leads connectors and acces...

Страница 23: ...in the industry Valuable to those with little knowledge of oscilloscopes as well as the experienced technician or engineer who wishes to refresh their memory or explore new uses for oscilloscopes Download your FREE copy from our Web site www bkprecision com ...

Страница 24: ... 2000 B K Precision Corp 481 314 9 001 Printed in Taiwan 1031 Segovia Circle Placentia CA 92870 ...

Отзывы: