10. CH1 AC-GND-DC Switch. Three-position lever
switch with the following positions:
AC:
Channel 1 input signal is capacitively coupled; dc
component is blocked.
GND:
Opens signal path and grounds input to vertical
amplifier. This provides a zero-volt base line, the
position of which can be used as a reference when
performing dc measurements.
DC:
Direct coupling of channel 1 input signal; both ac
and dc components of signal produce vertical de-
flection.
11.
CH1 (X) Input Jack. Vertical input for channel 1.
X-axis input for X-Y operation.
12.
CH1 (X) VOLTS/DIV Control. Vertical attenuator
for channel 1. Provides step adjustment of vertical
sensitivity. When channel 1 VARiable control is set
to CAL, vertical sensitivity is calibrated in 10 steps
from 5 mV/div to 5 V/div in a 1-2-5 sequence. When
the X-Y mode of operation is selected, this control
provides step adjustment of X-axis sensitivity.
13.
CH1 VARiable/PULL X5 MAG Control:
VARiable:
Rotation provides vernier adjustment of channel 1
vertical sensitivity. In the fully-clockwise (CAL)
position, the vertical attenuator is calibrated. Coun-
terclockwise rotation decreases gain sensitivity. In
X-Y operation, this control becomes the vernier
X-axis sensitivity control.
PULL X5 MAG:
When pulled out, increases vertical sensitivity by a
factor of five. Effectively provides two extra sensi-
tivity settings: 2 mV/div and 1 mV/div. In X-Y
mode, increases X-sensitivity by a factor of five.
14.
CH1
POSition/PULL CHOP Control:
POSition:
Rotation adjusts vertical position of channel 1 trace.
PULL CHOP:
When this switch is pulled out in the dual-trace
mode, the channel 1 and channel 2 sweeps are
chopped and displayed simultaneously (normally
used at slower sweep speeds). When it is pushed in,
the two sweeps are alternately displayed, one after
the other (normally used at higher sweep speeds).
15.
CH2
POSition/PULL INVert Control:
POSition:
Rotation adjusts vertical position of channel 2 trace.
In X-Y operation, rotation adjusts vertical position
of X-Y display.
PULL INVert:
When pushed in, the polarity of the channel 2 signal
is normal. When pulled out, the polarity of the
channel 2 signal is reversed, thus inverting the
waveform.
16.
CH2 VOLTS/DIV Control. Vertical attenuator for
channel 2. Provides step adjustment of vertical sensi-
tivity. When channel 2 VARiable control is set to
CAL, vertical sensitivity is calibrated in 10 steps from
5 mV/div to 5 V/div in a 1-2-5 sequence. When the
X-Y mode of operation is selected, this control pro-
vides step adjustment of Y-axis sensitivity.
17.
CH2 VARiable/PULL X5 MAG Control:
VARiable:
Rotation provides vernier adjustment of channel 2
vertical sensitivity. In the fully-clockwise (CAL)
position, the vertical attenuator is calibrated. Coun-
terclockwise rotation decreases gain sensitivity. In
X-Y operation, this control becomes the vernier
Y-axis sensitivity control.
PULL X5 MAG:
When pulled out, increases vertical sensitivity by a
factor of five. Effectively provides two extra sensi-
tivity settings: 2 mV/div and 1 mV/div. In X-Y
mode, increases Y-sensitivity by a factor of five.
18.
CH2 (Y) Input Jack. Vertical input for channel 2.
Y-axis input for X-Y operation.
19. CH2 AC-GND-DC Switch. Three-position lever
switch with the following positions:
AC:
Channel 2 input signal is capacitively coupled; dc
component is blocked.
GND:
Opens signal path and grounds input to vertical
amplifier. This provides a zero-volt base line, the
position of which can be used as a reference when
performing dc measurements.
DC:
Direct coupling of channel 2 input signal; both ac
and dc components of signal produce vertical de-
flection.
CONTROLS AND INDICATORS