background image

Trigger SOURCE Switch

The trigger SOURCE switch (CH 1CH 2, etc.) selects

the signal to be used as the sync trigger.

1.

If the SOURCE switch is set to CH 1 (or CH 2) the

channel 1 (or channel 2) signal becomes the trigger
source regardless of the VERTICAL MODE selec-
tion. CH 1, or CH 2 are often used as the trigger source
for phase or timing comparison measurements.

2.

By setting the SOURCE switch to ALT, alternating

triggering mode is activated. In this mode, the trigger
source alternates between CH 1 and CH 2 with each
sweep. This is convenient for checking amplitudes,
waveshape, or waveform period measurements, and
even permits simultaneous observation of two wave-
forms which are not related in frequency or period.
However, this setting is not suitable for phase or timing
comparison measurements. For such measurements,
both traces must be triggered by the same sync signal.
Alternate triggering can only be used in dual-trace
mode (VERT MODE set to DUAL), and with alter-
nate sweep only (PULL CHOP not engaged).

3.

In the EXT position of the COUPLING switch, the

signal applied to the EXT TRIG jack becomes the
trigger source. This signal must have a timing relation-
ship to the displayed waveforms for a synchronized
display.

4.

In the LINE position of the COUPLING switch,

triggering is derived from the input line voltage
(50/60 Hz) and the trigger SOURCE switch is dis-
abled. This is useful for measurements that are related
to line frequency.

TRIG LEVEL/PULL (–) SLOPE Control

(Refer to Fig. 2)

A sweep trigger is developed when the trigger source

signal crosses a preset threshold level. Rotation of the TRIG
LEVEL 
control varies the threshold level. In the direction
(clockwise), the triggering threshold shifts to a more posi-
tive value, and in the

direction (counterclockwise), the

triggering threshold shifts to a more negative value. When
the control is centered, the threshold level is set at the
approximate average of the signal used as the triggering
source. Proper adjustment of this control usually synchro-
nizes the display.

The TRIG LEVEL control adjusts the start of the sweep

to almost any desired point on a waveform. On sine wave
signals, the phase at which sweep begins is variable. Note
that  if the TRIG LEVEL control is rotated toward its
extreme or

setting, no sweep will be developed in the

normal trigger mode because the triggering threshold ex-
ceeds the peak amplitude of the sync signal.

When the PULL (–) SLOPE control is set to the (“in”)

position, the sweep is developed from the trigger source
waveform as it crosses a threshold level in a positive-going
direction. When the PULL (–) SLOPE control is set to the

(“out”) position, a sweep trigger is developed from the

trigger source waveform as it crosses the threshold level in
a negative-going direction.

MAIN TIME BASE Control

Set the Main Time Base TIME/DIV control to display

the desired number of cycles of the waveform. If there are
too many cycles displayed for good resolution, switch to a
faster sweep time. If only a line is displayed, try a slower
sweep time. When the sweep time is faster than the wave-
form being observed, only part of it will be displayed, which
may appear as a straight line for a square wave or pulse
waveform.

HOLDOFF Control

(Refer to Fig. 3)

A “holdoff” period occurs immediately after the comple-

tion of each sweep, and is a period during which triggering
of the next sweep is inhibited. The normal holdoff period
varies with sweep rate, but is adequate to assure complete
retrace and stabilization before the next sweep trigger is

Slope “–” Range

Slope “+” Range

Level

+

Fig. 2.

Function of Slope and Level Controls.

OPERATING INSTRUCTIONS

A. Holdoff not used

B. Holdoff used

Fig. 3.

Use of HOLDOFF Control.

14

Содержание 2190B

Страница 1: ...100 MHz DUAL TRACE OSCILLOSCOPE BK PRECISION INSTRUCTION MANUAL MODEL 2190B BK PRECISION ...

Страница 2: ...et 4 Use the time proven one hand in the pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return path 5 When using a probe touch only the insulated portion Never touch the exposed tip portion 6 When testing ac powered equipment remember that ac line voltage is usually present on some power input c...

Страница 3: ...l for further precautionary information This symbol appears in the manual where the corresponding information is given Instruction Manual for Model 2190B 100 MHz Dual Trace Oscilloscope 2000 B K Precision Corp 1031 Segovia Circle Placentia CA 92870 BK PRECISION ...

Страница 4: ...ing Procedure 12 Single Trace Display 12 Dual Trace Display 12 Triggering 13 Magnified Sweep Operation 15 OPERATING INSTRUCTIONS Continued X Y Operation 15 Video Signal Observation 15 Delayed Sweep Operation 15 Applications Guidebook 16 MAINTENANCE 17 Fuse Replacement 17 Line Voltage Selection 17 Periodic Adjustments 17 Calibration Check 18 Instrument Repair Service 18 APPENDIX 19 Important Consid...

Страница 5: ... with X10 MAG assures high frequencies and short duration pulses are displayed with high resolution VERTICAL FEATURES High Sensitivity 5 mV div sensitivity for full bandwidth High sen sitivity 1 mV div and 2 mV div using PULL X5 gain control Calibrated Voltage Measurements Accurate voltage measurements 3 on 10 cali brated ranges from 5 mV div to 5 V div Vertical gain fully adjustable between calib...

Страница 6: ...its stable observation of complex pulse trains OTHER FEATURES Bandwidth Limiter Selectable full bandwidth of 100 mHz or limited band width of 20 MHz to filter high frequency noise from low frequency waveforms X Y Operation Channel 1 can be applied as horizontal deflection X axis while channel 2 provides vertical deflection Y axis Built in Probe Adjust Square Wave A 2 V p p 1 kHz square wave genera...

Страница 7: ...proximately 500 kHz Polarity Reversal CH 2 invert Maximum Input Voltage 400 V dc ac peak HORIZONTAL AMPLIFIER Input through channel 1 input X Y mode CH 1 X axis CH 2 Y axis Sensitivity Same as vertical channel 2 Input Impedance Same as vertical channel 2 Frequency Response DC to 2 MHz 3 dB X Y Phase Difference 3 or less at 100 kHz Maximum Input Voltage Same as vertical channel 1 SWEEP SYSTEM Opera...

Страница 8: ...nominal CH 2 Y Output Output Voltage 50 mV div nominal into 50 ohm load Output Impedance Approximately 50 ohms Frequency Response 20 Hz to 100 MHz 3 dB Intensity Modulation Input Signal TTL level intensity increasing with more positive levels decreased intensity with more negative levels Input Impedance Approximately 50 kΩ Usable Frequency Range DC to 5 MHz Maximum Input Voltage 30 V dc ac peak Po...

Страница 9: ...ical calibration VERTICAL CONTROLS 9 VERTical MODE Switch Selects vertical display mode Four position lever switch with the following positions CH1 Displays the channel 1 signal by itself CH2 X Y CH2 displays the channel 2 signal by itself X Y used in conjunction with the X Y control and Trigger SOURCE switch to enable X Y display mode DUAL Displays the channel 1 and channel 2 signals simul taneou...

Страница 10: ...hen it is pushed in the two sweeps are alternately displayed one after the other normally used at higher sweep speeds 15 CH2 POSition PULL INVert Control POSition Rotation adjusts vertical position of channel 2 trace In X Y operation rotation adjusts vertical position of X Y display PULL INVert When pushed in the polarity of the channel 2 signal is normal When pulled out the polarity of the channe...

Страница 11: ...e delayed sweep X Y Used with the VERTical MODE switch and Trig ger SOURCE switch to select X Y operating mode The channel 1 input becomes the X axis and the channel 2 input becomes the Y axis Trigger source and coupling are disabled in this mode TRIGGERING CONTROLS 26 HOLDOFF PULL NORM TRIG Control HOLDOFF Rotation adjusts holdoff time trigger inhibit period beyond sweep duration When control is ...

Страница 12: ...ts a positive going slope and the out position selectsanegative going slopeas triggeringpoint for main sweep 30 EXTernal TRIGger Jack External trigger input for single and dual trace operation REAR PANEL CONTROLS not shown 31 Fuse Holder Line Voltage Selector Contains fuse and selects line voltage 32 Power Cord Receptacle 33 CH 2 Y SIGNAL OUTPUT Jack Output terminal where sample of channel 2 signa...

Страница 13: ...veform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never allow a small spot of high brilliance to remain stationary on the screen for more than a few seconds The screen may become permanently burned A spot will occur when the scope is set up for X Y operation and no signal is applied Either reduce the intensity so the spot i...

Страница 14: ...in Time Base control to 1 mS div Sweep Mode switch to MAIN 2 Press the red POWER pushbutton 3 A trace should appear on the CRT Adjust the trace brightness with the INTENSITY control and the trace sharpness with the FOCUS control NOTE When viewing audio and low frequency waveforms below 20 MHz a better wave form may be obtained by engaging the 20 MHz 100 MHz pushbutton This limits the bandwidth to ...

Страница 15: ...tain a stable jitter free dis play in single trace or dual trace operation The proper settings depend upon the type of waveforms being observed and the type of measurement desired An explanation of the various controls which affect synchronization is given to help you select the proper setting over a wide range of conditions AUTO or NORM Triggering 1 In the AUTO triggering mode PULL NORM TRIG swit...

Страница 16: ... direction counterclockwise the triggering threshold shifts to a more negative value When the control is centered the threshold level is set at the approximate average of the signal used as the triggering source Proper adjustment of this control usually synchro nizes the display The TRIG LEVEL control adjusts the start of the sweep to almost any desired point on a waveform On sine wave signals the...

Страница 17: ...he Trigger Source and VERTical MODE switches to X Y 2 In this mode channel 1 becomes the X axis input and channel 2 becomes the Y axis input The X and Y positions are now adjusted using the POSition and the channel 2 POSition controls respectively 3 Adjust the amount of vertical Y axis deflection with the CH 2 VOLTS DIV and VARIABLE controls 4 Adjust the amount of horizontal X axis deflection with...

Страница 18: ...ase control Because of this the oscilloscope automat ically prevents electrically the DELAY Time Base from being set to a slower sweep speed than the MAIN Time Base For example if the MAIN TimeBase is set to 0 1 ms div the slowest possibleDELAY Time Base sweep speed is also 0 1 ms div even if the control is set slower APPLICATIONS GUIDEBOOK B K Precision offers a Guidebook to Oscilloscopes which d...

Страница 19: ...le Pry fuseholder away from receptacle 3 When reinstalling fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To select the desired line voltage simply insert the fuse and fuse holder so that the appropriate voltage is pointed to by the arrow Be sure to use the proper value fuse see label on rear panel PERIOD...

Страница 20: ...d only as a general check of calibration accuracy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is required for calibration adjustments The CAL signal should not be used as a time base standard INSTRUMENT REPAIR SERVICE Because of the specialized skills and test equipment re quired for instrument repair...

Страница 21: ...p speed and using X10 magnification For the Model 2190B this sweep rate is 2 ns div A rise time of less than about five divisions at this sweep speed should be calculated Calculated Measurements For observed rise times of less than 11 ns the pulse rise time should be caluclated to eliminate the error introduced by the cascaded oscilloscope rise time Calculate pulse rise time as follows Tpulse Limi...

Страница 22: ...amages so the above limitation or exclusion may not apply to you This warranty gives you specific rights and you may also have other rights which vary from state to state SERVICE INFORMATION Warranty Service Please return the product in the original packaging with dated proof of purchase to the address below Clearly state in writing the performance problem and return any leads connectors and acces...

Страница 23: ...in the industry Valuable to those with little knowledge of oscilloscopes as well as the experienced technician or engineer who wishes to refresh their memory or explore new uses for oscilloscopes Download your FREE copy from our Web site www bkprecision com ...

Страница 24: ... 2000 B K Precision Corp 481 314 9 001 Printed in Taiwan 1031 Segovia Circle Placentia CA 92870 ...

Отзывы: