Trigger Coupling:
AUTO:
Sweep free-runs in absence of
suitable trigger signal.
NORM:
Sweep triggered only by adequate
trigger signal.
TV-V:
Video vertical sync pulses are
selected. Also usable for high
frequency reject.
TV-H:
Video horizontal sync pulses are
selected. Also usable for low
frequency reject.
Trigger Sensitivity:
AC:
1.5 div (internal)
≥
0.5 Vp-p (external)
30 Hz – 110 MHz
TV-V:
1.0 div (internal)
≥
0.5 Vp-p (external)
20 Hz – 30 kHz)
TV-H:
1.0 div (internal)
≥
0.5 Vp-p (external)
3 kHz – 100 kHz
Maximum External Trigger Voltage: 300 V (dc + ac
peak).
OTHER SPECIFICATIONS
Cal/Probe Compensation Voltage: 2 V p-p
±
3% square
wave, 1 kHz nominal.
CH 2 (Y) Output:
Output Voltage: 50 mV/div (nominal into 50 ohm
load).
Output Impedance: Approximately 50 ohms.
Frequency Response: 20 Hz to 100 MHz,
−
3 dB.
Intensity Modulation
Input Signal: TTL level, intensity increasing with
more positive levels, decreased intensity with more
negative levels.
Input Impedance: Approximately 50 k
Ω
.
Usable Frequency Range: DC to 5 MHz.
Maximum Input Voltage: 30 V (dc + ac peak).
Power Requirements: 100–130 VAC or 200–260 VAC,
50/60 Hz, 55 watts.
Dimensions (H
3
W
3
D):
5.2
″
3
12.8
″
3
15.7
″
(132
3
324
3
398 mm).
Weight: 18.8 lbs (8.5 kg).
Environment:
Within Specified Accuracy: +10° to +35° C, 10–80%
relative humidity.
Full Operation: 0° to +50° C, 10–80% relative
humidity.
Storage:
−
30° to +70° C, 10–90% relative humidity.
ACCESSORIES SUPPLIED:
Two Switchable X1/X10 Probes.
Instruction Manual.
AC Line Cord.
SPECIFICATIONS
6