Analyzer Control
The following SCPI commands control the generator outputs (see also page 19).
Note that even though the sampler configuration commands (be it PWM or NRZ) are located in a
sub-tree of each independent analyzer input, there is only one global configuration for all PWM
samplers and one global configuration for NRZ samplers. Therefore, changing the configuration (e.g.
the PWM data rate) for one analyzer input will actually affect all analyzer inputs.
:ANAlyzer:COUNt?
Syntax:
:ANAlyzer:COUNt?
Returns the total number of available analyzer inputs.
Example query:
ANA:COUN?
Example response:
2
:ANAlyzer#:IDENtifier?
Syntax:
:ANAlyzer#:IDENtifier?
Returns the unique identifier for this trigger input. This identifier is required when defining events
(see page 59).
Example query:
ANA0:IDEN?
Example response:
"ANALYZER0"
:ANAlyzer#:TERMinated
Syntax:
:ANAlyzer#:TERMinated <bool>
Specifies whether the relay that connects the termination resistors to ground is enabled or not.
Note that in single-ended mode (see
#ANAlyzer#:MODE
command on page 50), the complement
input must be left unconnected. See also section “Analyzer Module” on page 19.
Example command:
ANA0:TERM ON
:ANAlyzer#:TERMinated?
Syntax:
:ANAlyzer#:TERMinated?
Returns the current state of the termination relay.
Example query:
ANA0:TERM?
Example response:
1